Operator's Manual
174
ISSUED: October 2003
WR6K-OM-E Rev A
discriminate levels
from noise in data.
Available with
JTA2 and XMAP
options.
Edge@level
Number of edges in waveform.
Reference levels
and edge-
transition polarity
can be selected.
Hysteresis
argument used to
discriminate levels
from noise in data.
Available with
JTA2 and XMAP
options.
Excel
Performs measurements in
Excel by transferring one or two
waveforms and reading the
resulting parameter value.
Available with
XMAP option.
Excel must be
loaded on the
instrument.
Fall time
Fall time: Duration of falling
edge from 90-10%.
Threshold arguments specify
two vertical values on each
edge used to compute fall time.
Formulas for upper and lower
values:
lower = lower thresh. x
amp/100 + base
upper = upper thresh. x
amp/100 + base
Time at upper threshold
minus
Time at lower threshold
averaged over each falling
edge
On signals not
having two major
levels (triangle or
saw-tooth waves,
for example), top
and base can
default to
maximum and
minimum, giving,
however, less
predictable
results.
Standard
parameter.
Fall 80-20%
Fall 80-20%: Duration of pulse
waveform's falling transition
Average duration of falling
80-20% transition
On signals not
having two major
Summary of Contents for 6000 SERIES
Page 65: ...WR6K OM E Rev A ISSUED October 2003 63 Proper Orientation of Drive ...
Page 131: ...WR6K OM E Rev A ISSUED October 2003 129 Example ...
Page 141: ...WR6K OM E Rev A ISSUED October 2003 139 ...
Page 236: ...Operator s Manual 234 ISSUED October 2003 WR6K OM E Rev A Example 6 ...
Page 261: ...WR6K OM E Rev A ISSUED October 2003 259 ...
Page 286: ...Operator s Manual 284 ISSUED October 2003 WR6K OM E Rev A ...
Page 304: ...Operator s Manual 302 ISSUED October 2003 WR6K OM E Rev A ...
Page 307: ...WR6K OM E Rev A ISSUED October 2003 305 Convolve two signals ...
Page 348: ...Operator s Manual 346 ISSUED October 2003 WR6K OM E Rev A ...
Page 402: ...Operator s Manual 400 ISSUED October 2003 WR6K OM E Rev A ...