5.
Touch inside the
Attenuation
field and select an attenuation factor from the pop-up
menu.
6.
Touch the
Current
tab, then touch inside the current
Source
field and select a channel,
memory, or math trace from the pop-up menu.
7.
Touch inside the
Fine Adjust DC Level
field and enter a value, using the pop-up keypad.
8.
Touch inside the
Current probe type
field and make a selection from the pop-up menu.
The display should appear similar to this:
Note:
It is important to set V/div to allow the largest expected voltage excursion to remain on
screen during all conditions of the test. If the voltage signal goes off screen during the test,
erroneous results will be obtained.
9.
It is good practice to balance the DC offset of the voltage probe at this point. Press the
DA1855A's Autozero button to autobalance the amplifier.
Note:
The channel selected as the Current Input channel will be assigned Ampere units even if a
voltage or non–ProBus-compatible current probe is used.
10.
It is also good practice to degauss the current probe at this point. Remove the current
probe from the circuit under test, close the probe and press Degauss.
11.
Reconnect the current probe to the circuit under test.
Summary of Contents for 6000 SERIES
Page 65: ...WR6K OM E Rev A ISSUED October 2003 63 Proper Orientation of Drive ...
Page 131: ...WR6K OM E Rev A ISSUED October 2003 129 Example ...
Page 141: ...WR6K OM E Rev A ISSUED October 2003 139 ...
Page 236: ...Operator s Manual 234 ISSUED October 2003 WR6K OM E Rev A Example 6 ...
Page 261: ...WR6K OM E Rev A ISSUED October 2003 259 ...
Page 286: ...Operator s Manual 284 ISSUED October 2003 WR6K OM E Rev A ...
Page 304: ...Operator s Manual 302 ISSUED October 2003 WR6K OM E Rev A ...
Page 307: ...WR6K OM E Rev A ISSUED October 2003 305 Convolve two signals ...
Page 348: ...Operator s Manual 346 ISSUED October 2003 WR6K OM E Rev A ...
Page 402: ...Operator s Manual 400 ISSUED October 2003 WR6K OM E Rev A ...