Operator's Manual
254
ISSUED: October 2003
WR6K-OM-E Rev A
The next figure shows a part of the Excel workbook.
Here we see the gated waveform that has been created in Excel. The Mean parameter during the
region of interest (ROI) is placed in cell H3.
How Does this Work?
The amplitude of the signal is about 0.3 volts, and the screen height is 0.4 volts, as derived from
cells F7 and Fx. A threshold value for amplitude was calculated by placing 0.5 * (Fy – Fx) in cell
A4.
Remember that in the instrument the sources were defined to be A10 and B10. This means that
the first point on the waveform will be read into A10, and, since the waveform has 500 points, the
last point will be read into A510. The same holds true for F2 and column B, since F2 is assigned
as Source2, and data is defined to write into column B starting with cell B10.
To create the gating function in column C, the cell C10 was given the following formula:
IF ( ( B10 – B9) > $A$4, 1 – C9, C9). This was copied down the column. Column D, the output
column, is simply A * C.
The output was defined as cell H3.
The required mean in cell H3 is given by SUM (D10 : D509) / SUM (C10 : C509), for a 500 point
waveform.
Summary of Contents for 6000 SERIES
Page 65: ...WR6K OM E Rev A ISSUED October 2003 63 Proper Orientation of Drive ...
Page 131: ...WR6K OM E Rev A ISSUED October 2003 129 Example ...
Page 141: ...WR6K OM E Rev A ISSUED October 2003 139 ...
Page 236: ...Operator s Manual 234 ISSUED October 2003 WR6K OM E Rev A Example 6 ...
Page 261: ...WR6K OM E Rev A ISSUED October 2003 259 ...
Page 286: ...Operator s Manual 284 ISSUED October 2003 WR6K OM E Rev A ...
Page 304: ...Operator s Manual 302 ISSUED October 2003 WR6K OM E Rev A ...
Page 307: ...WR6K OM E Rev A ISSUED October 2003 305 Convolve two signals ...
Page 348: ...Operator s Manual 346 ISSUED October 2003 WR6K OM E Rev A ...
Page 402: ...Operator s Manual 400 ISSUED October 2003 WR6K OM E Rev A ...