Operator's Manual
340
ISSUED: October 2003
WR6K-OM-E Rev A
8.
A second setup dialog opens to the right of the main with more setup options. The
options offered depend on the parameter you chose, but all include
Level is
,
Percent
Level
,
Slope
, and
Hysteresis
. A
Find Level
button is also provided in this mini-dialog.
Option Field
Settings
Level Is
Percent Absolute
Percent Level
0 to 100%
Slope Positive
Negative
Hysteresis
0 div to 10 div
Note:
The
Hysteresis
selection imposes a limit above and below the Level, which precludes
measurements of noise or other perturbations within this band. The width of the band is specified
in milli-divisions.
Guidelines for Use
1. Hysteresis must be larger than the maximum noise spike you want to ignore.
2. The largest value of hysteresis usable is less than the distance from the level to the closest
extreme value of the waveform.
3. Unless you know the largest noise and closest extreme level that will ever occur on any cycle,
leave some margin on both sides of the level.
Summary of Contents for 6000 SERIES
Page 65: ...WR6K OM E Rev A ISSUED October 2003 63 Proper Orientation of Drive ...
Page 131: ...WR6K OM E Rev A ISSUED October 2003 129 Example ...
Page 141: ...WR6K OM E Rev A ISSUED October 2003 139 ...
Page 236: ...Operator s Manual 234 ISSUED October 2003 WR6K OM E Rev A Example 6 ...
Page 261: ...WR6K OM E Rev A ISSUED October 2003 259 ...
Page 286: ...Operator s Manual 284 ISSUED October 2003 WR6K OM E Rev A ...
Page 304: ...Operator s Manual 302 ISSUED October 2003 WR6K OM E Rev A ...
Page 307: ...WR6K OM E Rev A ISSUED October 2003 305 Convolve two signals ...
Page 348: ...Operator s Manual 346 ISSUED October 2003 WR6K OM E Rev A ...
Page 402: ...Operator s Manual 400 ISSUED October 2003 WR6K OM E Rev A ...