Operator's Manual
374
ISSUED: October 2003
WR6K-OM-E Rev A
Instantaneous Power Measurement – Steady State
Note:
The previous sections should be completed before you make these measurements.
Selecting
Instantaneous Power
from the
Analysis
menu brings up the following display:
•
The deskewed current waveform (I) and the voltage waveform (V) are displayed in the first
grid.
•
The instantaneous power waveform (Pwr) is displayed in the second grid.
1.
Fine Adjust DC Level
for both voltage and current are provided to allow you to
compensate for any residual DC offset that the voltage and current probes may exhibit.
Because the magnitude of residual DC offset error is usually unknown, use of these
adjustments requires that you know something about the waveforms.
2.
Press the
Fine
button, then turn the
Adjust
knob to set the DC offset level.
3.
For a detailed explanation of how to adjust the DC offset level, see Fine DC Level
Adjustment.
Summary of Contents for 6000 SERIES
Page 65: ...WR6K OM E Rev A ISSUED October 2003 63 Proper Orientation of Drive ...
Page 131: ...WR6K OM E Rev A ISSUED October 2003 129 Example ...
Page 141: ...WR6K OM E Rev A ISSUED October 2003 139 ...
Page 236: ...Operator s Manual 234 ISSUED October 2003 WR6K OM E Rev A Example 6 ...
Page 261: ...WR6K OM E Rev A ISSUED October 2003 259 ...
Page 286: ...Operator s Manual 284 ISSUED October 2003 WR6K OM E Rev A ...
Page 304: ...Operator s Manual 302 ISSUED October 2003 WR6K OM E Rev A ...
Page 307: ...WR6K OM E Rev A ISSUED October 2003 305 Convolve two signals ...
Page 348: ...Operator s Manual 346 ISSUED October 2003 WR6K OM E Rev A ...
Page 402: ...Operator s Manual 400 ISSUED October 2003 WR6K OM E Rev A ...