16. EXTERNAL INTERFACE
168
● Tally Display
A tally is displayed in green when pin 13 (TALLY) is connected to ground.
A tally can be displayed only when the measurement mode is set to picture, CINELITE, or
CINEZONE.
Figure 16-4 Tally display
● Alarm output
The LV 5333 outputs an alarm from pin 14 (ALARM) when any of the conditions below is
true.
To select the alarm output polarity, press REMOTE on the STATUS menu.
• When an error occurs in the item that you have set to INT&RMT or ALL ON using ERROR
DETECT on the STATUS menu
• When the temperature inside the instrument exceeds 80°C
• When a fan error occurs
Reference REMOTE → 5.4.1, “Configuring the Remote Connector”
ERROR DETECT →13.6.3, “Configuring Error Detection Settings”