159
14. AUDIO DISPLAY
F•4
AUDIO → F•4 LEVEL METER → F•2 RANGE: PEAK60dB / PEAK90dB / AVERAGE
14.4 Selecting the Audio Output Channels
To select the audio output channels, follow the procedure below.
You can select the audio output channels from those selected with SDI GROUP.
Reference
SDI Group → 14.3, “Selecting the Channels to Measure”
Procedure
AUDIO → F•3 AES/EBU OUTPUT: CH1/2 / CH3/4 / CH5/6 / CH7/8 / CH9/10 / CH11/12 /
CH13/14 / CH15/16
14.5 Configuring the Meter Settings
To configure meter settings, press
LEVEL METER on the AUDIO menu.
→
LEVEL METER →
Figure 14-_5 LEVEL METER menu
14.5.1
Selecting the Reference Level
To select the meter reference level, follow the procedure below.
Levels that exceed the reference level are displayed in red, and levels that are less than the
reference level are displayed in green.
Procedure
14.5.2
Selecting the Range
To set the meter's range, follow the procedure below.
Procedure
Settings
PEAK60dB: A meter with a peak marker and a range from -60 to 0 dB
PEAK90dB: A meter with a peak marker and a range from -90 to 0 dB
AVERAGE: A meter with a peak marker, a reference level at 0 dB, and a range from -20 to
3 dB
The following table shows meter response models.
Table 14-1 Meter response models
AUDIO → F•4 LEVEL METER → F•1 REF LEVEL:
-
20dB /
-
18dB /
-
12dB /
-
9dB
F•4
AUDIO
RANGE
delay time (*1)
return time (*2)
PEAK60dB
Instantaneous
1.7 sec
PEAK90dB
Instantaneous
1.7 sec
AVERAGE
0.3 sec
0.3 sec