LTI TruPoint 200h User’s Manual • Section 7 - Targeting Menu
Page 58
Copyright (c)[2020] Laser Technology, Inc. All Rights Reserved. Unauthorized duplication, in whole or in part is strictly prohibited.
To select:
1.
Short press the
button. Use use the
and
buttons to the to highlight the Range
Gate Icon
. Press the
button.
2.
Use the
and
buttons to highlight either gate. To set the value, press the
button
on the desired gate. Edit right/left arrows will be displayed. Increase or decrease uses only
the
or
buttons no matter how the unit positioned.
•
It is orientation-independent and the
button is not active during
this application.
•
Short press the arrow buttons to change the value by 1
•
Press-and-hold the arrow buttons to change the value faster.
3.
Press the
button to accept/set the value and the S/L Gate icon will be highlighted.
4.
Continue/repeat steps to set the other gate if necessary.
•
Short press the right soft key
to set and return to the main measurement screen.
•
Short press the Left Soft Key
to cancel and exit to Targeting Menu screen.
If a gate is set, the gate icon will be displayed in the lower status bar section of
the main measurement screen.
Limits:
•
Short and Far gate limit: 1 - 500 m (1 - 1,500 ft)
•
EXAMPLE: The gate window is restricted to a width of 1.5 meters (5
feet). If you set the gate window too narrow, the instrument
automatically adjusts it by resetting either the short or long gate,
taking the most recent value entered as the standard to recalculate the
other by. If the short gate is already set at 10 feet, for example, and
then you set the long gate to 12 feet, the instrument resets the short
gate to 7 feet.
•
“DISABLED” or ---m if gate is disabled.
•
If the unit does not acquire a measurement within the gate window,
it will show an 'E01' message (No Target Acquired) and stop
measuring.The unit will not keep trying to get a measurement
at this point.
**Draft B**