SoundAdvisor Model 831C
Powering the 831C SLM
3-6
Step 4
Select
Save
, then click
Yes
to confirm your choice.
TAKE NOTE
Batteries charge more efficiently when the meter is powered off. For
details about 831C button LED indicators, see “Understanding Charging
Step 5
Power the meter OFF.
Step 6
Connect the meter (with the USB cable) to the included power cable
(PSA029), and plug this in to a wall outlet or other external power
source.
TAKE NOTE
If you’re using a power cable other than the one supplied with the
meter, use a 28 AWG or larger wire for proper charging.
Understanding Charging Status LED Indicators
The charge status indicated by the green LED behind the
Power
button
is as follows:
• Solid green: Charging
• No green light: Not charging
• Green light wink: Maintaining charge
• Green light fast blink (two times per second): Charger fault
• Green light fastest blink (three times per second): Meter is
powering up or shutting down
Resolving a “Charging Stopped” Session Log Entry
To resolve a “Charging Stopped” Session Log entry, check the battery
placement, and make sure the correct
Battery Type
is selected (
Tools
System Properties
).
Resolving an “Incorrect Battery Type” Session Log Entry
Infrequently, when the battery becomes too old to charge, the meter
may register the battery as the incorrect battery type. This can also
occur when newer NiMH batteries are fully discharged. In this case,
charge the batteries using an external battery charger, then install
them in the meter.
You may also see this error when the batteries become too hot or too
cold. Optimal battery temperature is 0° C–45° C.
3.5.2 Using External Power
We recommend maintaining the internal battery The 831C utilizes the
following external power sources:
• USB port from a computer
• USB port from PSA029 power supply
• I/O connector from an external +10.8 to +25 Volt DC mains
power source.