442700/A
255
Testing the noise conditions and the sonar
functionality
Topics
Checking the status of the transceiver boards, page 255
Checking the transducer by means of the Diagnostics functionality, page 257
Measuring the noise in passive mode, page 261
Making a noise/speed curve to determine vessel noise, page 263
Measuring flow noise, page 267
Checking the status of the transceiver boards
The purpose of the
Diagnostics
dialog box is to allow on-line testing and verification of the
hardware components in the CS90 system. The
Transceiver
page offers key information
about each transceiver board in the Transceiver Unit.
Prerequisites
Neither tools nor instruments are required.
• The CS90 system is turned on and operates normally.
• The CS90 system is in
Normal
mode, but
TX Power
is set to
Off
to prevent transmissions.
• The transducer is retracted to its upper position.
• The vessel is berthed or at sea. We recommend that the test is done with the vessel at sea.
Caution
You must never set the CS90 system to "ping" unless the transducer is submerged in water.
The transducer may be damaged if it transmits in open air.
Context
By means of the
Diagnostics
functionality, you can easily determine if the hardware
components in the CS90 system are operational. And most important, you can make sure
that all the transceivers channels and transducer elements are functional. Each button
provides a small colour coded indicator.
•
No indicator
: Status is OK. No actions are necessary.
Summary of Contents for Simrad CS90
Page 1: ...kongsberg com simrad Simrad CS90 REFERENCE MANUAL ...
Page 2: ......
Page 627: ...442700 A 625 Related topics Concept descriptions page 622 Concept descriptions ...
Page 683: ......
Page 684: ... 2021 Kongsberg Maritime ISBN 978 82 8066 213 2 ...
Page 685: ......
Page 686: ...Reference Manual Simrad CS90 ...