![Knick Stratos Pro A201 PH User Manual Download Page 99](http://html.mh-extra.com/html/knick/stratos-pro-a201-ph/stratos-pro-a201-ph_user-manual_1980194099.webp)
99
Step
Action/Display
Remark
Select HOLD
mode
Press
menu
key to call
the selection menu,
select HOLD using the
keys, press
enter
to confirm.
Now the device is in HOLD
mode. The HOLD mode can also
be activated externally via the
HOLD input. During HOLD the
output current is frozen at its last
value or set to a fixed value.
Disconnect old
sensor,
connect new
sensor.
Temporary messages are dis-
played during the replacement
but neither output to the alarm
contact nor entered in the log-
book.
Wait until the
sensor data are
displayed.
MEMO SENS
Check sensor
data
MEMO SENS
View sensor informa-
tion using
keys,
press
enter
to confirm.
You can view the sensor manu-
facturer and type, serial number,
and last calibration date.
Check
measured
values,
then
exit HOLD.
Hit
meas
key:
Return to the selec-
tion menu. Hold
meas
key depressed: Device
switches to measuring
mode.
The sensor replacement is
entered in the extended logbook
(TAN SW-A003).