Alarm
Manual activation of HOLD
The HOLD mode can be activated manually from the HOLD menu. This
allows checking or replacing a sensor, for example, without provoking
unintended reactions at the outputs.
Press
meas
key to return to selection menu.
Alarm
When an error has occurred,
Err xx
is displayed immediately.
Only after expiry of a user-defined delay time will the alarm be
registered and entered in the logbook.
During an alarm the display blinks, the display backlighting turns
red
.
Error messages can also be signaled by a 22 mA output current
(see Configuration).
2 sec after the failure event is corrected, the alarm status will be
deleted.
External activation of HOLD (TAN SW-A005)
The HOLD mode can be activated from outside by sending a signal to
the HOLD input (e.g. from the process control system).
HOLD inactive
0...2 V AC/DC
HOLD active
10...30 V AC/DC
Process control system
Power supply
12...24 V AC/DC
HOLD
input
10
11
Stratos Pro A201