Configuration
Configuration
Select
Default
Real-time clock (CLOCK)
CLK:
FORMAT
24 h / 12 h
24 h
24 h
TIME hh/mm
00..23:00...59
12 h
TIME hh/mm
00...11 AM/PM: 00...59
DAY/MONTH
01...31/01...12
YEAR
2000...2099
Measuring points (TAG / GROUP)
TAG:
(Input in text line)
A...Z, 0...9, – + < > ? / @
GROUP:
(Input in text line)
0000...9999
Display backlighting (DISPLAY)
DSP:
BACKLIGHT
On, Off
On
*) Hysteresis fixed at 5% of threshold value
Control input (CNTR_IN)
IN:
CONTROL
Parameter‑set
switchover (PARSET)
or
flow measurement
(FLOW)
FLOW
FLOW ADJUST
0 ... 20000
pulses/liter
12000
pulses/liter
Alarm (ALARM)
ALA:
DELAYTIME
0...600 SEC
0010 SEC
SENSOCHECK
ON/OFF
OFF
FLOW CNTR
**)
ON/OFF
OFF
ON
FLOW MIN
**)
0 ... 99.9 L/h
005.0 L/h
FLOW MAX
**)
0 ... 99.9 L/h
025.0 L/h
Parameter set selection (PARSET)
PAR:
Select fixed parameter set
(A) or switch between A/B
via control input or manu‑
ally in measuring mode
PARSET FIX A/
CNTR INPUT /
MANUAL
PARSET FIX A
Summary of Contents for Stratos Pro A201 PH