ORP Calibration
Display
Action
Remark
Select ORP calibration,
proceed with
enter
Remove the sen-
sor and temperature
probe, clean them, and
immerse them in the
redox buffer.
Display (3 sec)
Now the device is in
HOLD mode.
Enter setpoint value for
redox buffer.
Press
enter
to proceed.
The ORP delta value is
displayed (based on
25°C).
Sensoface is active.
Press
enter
to proceed.
To repeat calibration:
Select REPEAT.
To exit calibration:
Select MEAS, then press
enter
After end of calibra-
tion, the outputs re-
main in HOLD mode
for a short time.