![Knick Stratos Pro A201 PH User Manual Download Page 132](http://html.mh-extra.com/html/knick/stratos-pro-a201-ph/stratos-pro-a201-ph_user-manual_1980194132.webp)
132
Specifications
Diagnostic functions
Calibration data
Calibration date, zero, slope, response time
Device self-test
Display test, Automatic memory test (RAM, FLASH, EEPROM), module
test
Logbook (TAN)
100 events with date and time
Extended logbook (TAN)
Audit Trail: 200 events with date and time
Service functions
Sensor monitor
Display of direct sensor signals
Current source
Current specifiable for output 1 and 2 (04.00 ... 22.00 mA)
Passcodes
Assignment of passcodes for access to menus
Factory setting
Reset all parameters to factory settings
TAN
Enable optionally available add-on functions
Data retention
Parameters, calibration data, and logbook > 10 years (EEPROM)
Housing
Molded enclosure, glass fiber reinforced
Front unit material: PBT
Rear unit material: PC
Mounting
Wall, pipe/post or panel mounting
Color
Gray RAL 7001
Ingress protection
IP66/IP67/TYPE 4X outdoor (with pressure compensation) when the
device is closed
Flammability
UL 94 V-0 for external parts
Dimensions
148 mm x 148 mm
Control panel cutout
138 mm x 138 mm acc. to DIN 43 700
Weight
approx. 1200 kg (1.6 kg incl. accessories and packaging)
Cable glands
5 knockouts for M20 x 1.5 cable glands
2 of 5 knockouts for NPT ½” or rigid metallic conduit
Terminals
Screw terminals
for single or stranded wires 0.2… 2.5 mm
2
Tightening torque
0.5 … 0.6 Nm