Configuration
Menu item
Action
Choices
CIP / SIP
The following adjustments are possible for standard and ISFET
sensors:
Cleaning cycles
Select ON or OFF using
keys.
When switched on, the
cycles will be entered in
the extended logbook
(TAN SW‑A003) but will
not be counted.
Press
enter
to confirm.
Sterilization cycles
Select ON or OFF using
keys.
When switched on, the
cycles will be entered in
the extended logbook
(TAN SW‑A003) but will
not be counted.
Press
enter
to confirm.
Logging the cleaning and sterilization cycles with connected sensor
helps measuring the load on the sensor.
Suitable for biochemical applications
(process temp approx. 0 ... 50 °C / 32 ... 122 °F,
CIP temp > 55 °C / 131 °F,
SIP temp > 115 °C / 239 °F).
Note:
A CIP or SIP cycle is only entered into the extended logbook (TAN
SW‑A003) 2 hours after the start to ensure that the cycle is complete.
Configuration