Configuration
Menu item
Action
Choices
CIP / SIP
The following adjustments are possible for ISM sensors :
Cleaning cycle
counter
Select ON or OFF using
keys.
Press
enter
to confirm.
ON/
OFF
Cleaning cycles
Only with CIP COUNT ON:
Enter value using
keys.
Press
enter
to confirm.
0...9999 CYC
(
0000 CYC
)
Sterilization cycle
counter
Select ON or OFF using
keys.
Press
enter
to confirm.
ON/
OFF
Sterilization cycles
Only with CIP COUNT ON:
Enter value using
keys.
Press
enter
to confirm.
0...9999 CYC
(
0000 CYC
)
Logging the cleaning and sterilization cycles with connected sensor
helps measuring the load on the sensor.
Suitable for biochemical applications (process temp approx.
0 ... +50 °C / +32 ... +122 °F, CIP temp > +55 °C / +131 °F,
SIP temp > +115 °C / +239 °F).