![Knick Stratos Pro A201 PH User Manual Download Page 152](http://html.mh-extra.com/html/knick/stratos-pro-a201-ph/stratos-pro-a201-ph_user-manual_1980194152.webp)
152
Error
Info text
(is displayed in case of
fault when the Info key
is pressed)
Problem
Possible causes
ERR 05
CAL DATA
Error in cal data *
ERR 10
ORP RANGE
ORP display range limits
exceeded
< -1999 mV or > 1999 mV
ERR 11
PH RANGE
pH display range limits
exceeded
< -2 or > 16
ERR 12
MV RANGE
mV range
ERR 13
TEMPERATURE RANGE
Temperature range limits
exceeded
Connect the sensor, check the sensor
cable and replace if necessary, check
the sensor connection, adjust the
parameter settings.
ERR 15
SENSOCHECK GLASS-EL
Sensocheck glass
ERR 16
SENSOCHECK REF-EL
Sensocheck ref.
ERR 60
OUTPUT LOAD
Load error
Check the current loop, deactivate
unused current outputs.
ERR 61
OUTPUT 1 TOO LOW
Output current 1
< 3.8 mA
ERR 62
OUTPUT 1 TOO HIGH
Output current 1
> 20.5 mA
ERR 63
OUTPUT 2 TOO LOW
Output current 2
< 3.8 mA
ERR 64
OUTPUT 2 TOO HIGH
Output current 2
> 20.5 mA
Error Messages