120
Diagnostics
Display
Menu item
Display examples:
Displaying the currently measured values
(sensor monitor)
Select MONITOR using
, press
enter
to confirm.
Use the
keys to select the desired parameter
from the bottom line of the display: mV_PH mV_ORP
RTD R_GLASS R_REF I-INPUT (for digital sensors also:
OPERATION TIME SENSOR WEAR LIFETIME CIP SIP
AUTOCLAVE, for ISM sensors in addition:
ACT (adaptive calibration timer)
TTM (adaptive maintenance timer)
DLI (Dynamic Life Time Indicator). The selected
parameter is shown in the main display.
Press
meas
to return to measurement.
Display mV_pH
(for validation, sensor can be immersed in a calibra-
tion solution, for example, or the device is checked by
using a simulator)
Display of remaining dynamic lifetime
(only for digital sensors, however not for MEMOSENS)
Display of sensor operating time
(for digital sensors only)
Version
Display
of device type, software/hardware version
and
serial number
for all device components.
Use the
keys to switch between software and
hardware version. Press
enter
to proceed to next
device component.
Summary of Contents for Stratos Pro A201 PH