Configuration
Menu item
Action
Choices
Process variable
Select using
keys:
PH: pH value
ORP: Redox potential
RH: rH value
(with pH/ORP sensor)
TMP: Temperature
Press
enter
to confirm.
PH
/ORP/RH/TMP
Current start
Modify digit using
keys,
select next digit using
keys.
Press
enter
to confirm.
0 ... 42.5 rH
Current end
Enter value using
keys.
Press
enter
to confirm.
0 ... 42.5 rH
Assignment of measured values: Current start and current end
Output current
[pH]
20
4
14
0
20
4
7
5
[mA]
Example 1: Range pH 0...14
[mA]
Output current
7
5
5
Example 2: Range pH 5...7
Advantage: Higher resolution in
range of interest
[pH]