106
Manual Calibration with Buffer Entry
The MAN calibration mode and the type of temperature detection
are selected during
configuration
. For calibration with manual buffer
specification, you must enter the pH value of the buffer solution used
in the device for the proper temperature. Any desired buffer solution
can be used for calibration.
Display
Action
Remark
Select Calibration.
Press
enter
to proceed.
Ready for calibration.
Hourglass blinks.
Display (3 sec)
Now the device is in
HOLD mode.
Remove the sensor and
temperature probe,
clean them, and im-
merse them in the first
buffer solution.
Press
enter
to start.
When manual input
of temperature has
been configured, the
temp value in the
display blinks and
can be edited using
the arrow keys.
Enter the pH value of
your buffer solution for
the proper temperature.
While the “hourglass”
icon is blinking, the
sensor and temperature
probe remain in the
buffer solution.
The response time
of the sensor and
temperature probe
is considerably
reduced when the
sensor is first moved
about in the buffer
solution and then
held still.