46
Configuration
Configuration
Select
Default
Sensor (SENSOR)
SNS: ISM
ACT
(Adaptive
Calibration Timer)
OFF
AUTO
MAN
OFF
MAN ACT CYCLE 0...9999 DAY
0007 DAY
TTM
(Time to
Maintenance)
OFF
AUTO
MAN
OFF
MAN TTM CYCLE 0...9999 DAY
0030 DAY
Memosens,
Inducon,
ISM
CIP COUNT
ON/OFF
OFF
ON CIP CYCLES
0...9999 CYC
0025 CYC
SIP COUNT
ON/OFF
OFF
ON SIP CYCLES
0...9999 CYC
0025 CYC
AUTOCLAVE
ON/OFF
OFF
ON AC CYCLES
0...9999 CYC
0000 CYC
CHECK TAG/
GROUP
ON / OFF
OFF
Output 1 (OUT1)
BEGIN 4 mA
END 20 mA
BEGIN 4 mA
END 20 mA
BEGIN 4 mA
Select °C / °F at
“Sensor“
END 20 mA
BEGIN 4 mA
END 20 mA
FILTERTIME
0...120 SEC
0000 SEC
22mA FAIL
ON/OFF
OFF
22mA FACE
ON/OFF
OFF
HOLD MODE
LAST/FIX
LAST
FIX
HOLD‑FIX
04.00...22.00 mA
021.0 mA