129
-13- Kraft
2.00/4.00/7.00/9.00/11.00
-U1-
Specifiable buffer set with 2 buffer solutions
Zero offset
± 200 mV (ISFET only)
Max. calibration range
Asymmetry potential
±60 mV
Slope
80 ... 103 % (47.5 ... 61 mV/pH)
(possibly restricting notes from Sensoface)
ORP sensor standardization
*)
ORP calibration (zero offset)
Max. calibration range
–700 ... +700 ΔmV
Adaptive calibration timer
*)
Interval 0000 ... 9999 h
Sensocheck
Automatic monitoring of glass and reference electrode
(can be switched off)
Delay
Approx. 30 s
Sensoface
Provides information on the sensor condition
Evaluation of zero/slope, response time, calibration interval, wear,
Sensocheck; can be disabled
Temperature input
Pt100 / Pt1000 / NTC 30 kΩ / NTC 8,55 kΩ / Balco 3 kΩ
*)
2-wire connection, adjustable
Measuring range
Pt 100/Pt 1000
–20.0 ... 200.0 °C / –4 ... 392 °F
NTC 30 kΩ
–20.0 ... 150.0 °C / –4 ... 302 °F
NTC 8.55 kΩ (Mitsubishi) –10.0 ... 130.0 °C / 14 ... 266 °F
Balco 3 kΩ
–20.0 ... 130.0 °C / –4 ... 266 °F
Adjustment range
10 K
Resolution
0.1 °C / 0.1 °F
Measurement error
1,2,3)
< 0.5 K (± 1 K for Pt100; < 1 K for NTC > 100°C)
Temperature compensation
of process medium
Linear –19,99 ... +19.99 %/K (reference temp 25 °C)
Ultrapure water, user-defined table
ISM input
“One wire” interface for operation with ISM (digital sensors)
(6 V / Ri= approx. 1.2 kΩ)
Power output
for operating an ISFET adapter
+3 V / 0.5 mA
-3 V / 0.5 mA