Configuration
Menu item
Action
Choices
Select sensor type
Select sensor type using
keys.
Press
enter
to confirm.
Digital sensors:
MEMOSENS
INDUCON
ISM
Select type of temp
probe
(not for digital sensors)
Select type of tempera‑
ture probe using
keys.
Press
enter
to confirm.
1000 PT, 100 PT,
Balco 3 kΩ, 8.55 NTC,
30 NTC
Temperature unit
Select °C or °F using
keys.
Press
enter
to confirm.
Temp detection
during measurement
Select mode using
keys:
AUTO: Measured by
sensor
MAN: Direct input of
temperature, no measure‑
ment (see next step)
EXT: Temperature speci‑
fied via current input
(only if TAN E enabled)
Press
enter
to confirm.
(Manual temperature)
Modify digit using
keys,
select next digit using
keys.
Press
enter
to confirm.