Configuration
Menu item
Action
Choices
Temp compensation,
process medium
For pH measurement
only: Select temperature
compensation of the
process medium.
Linear: LIN
Ultrapure water:
PUREWTR
Table: USERTAB
Select using
key,
proceed using
enter
.
Temp compensation,
linear
Enter the linear tempera‑
ture compensation of the
process medium.
Enter value using
keys.
Press
enter
to confirm.
Temp compensation,
table
Confirm safety
prompt,
then enter values
(5 °C step size)
When you have selected
temperature compensa‑
tion via table (USERTAB),
you can enter values for
a TC table from 0 to 95 °C
in 5‑K steps.
The analyzer displays
temperature values in
5 °C steps. You must spec‑
ify the percent deviation
of the measured value
from each of these tem‑
perature values.
Intermediate values are
linearly interpolated.
TC compensation can be
defined separately for
parameter sets A and B.
0 ... 95 °C
(5 °C step size)