®
User Manual
Stratos Pro A201 PH
pH Measurement
Read before installation.
Keep for future use.
www.knick.de
Page 1: ... User Manual Stratos Pro A201 PH pH Measurement Read before installation Keep for future use www knick de ...
Page 2: ... the reader a basic understanding of safety It illustrates general hazards and gives strate gies on how to avoid them Safety Guide The external Safety Guide is designed to give the reader a basic understanding of safety It illustrates general hazards and suggests strategies on how to avoid them Warnings This document uses the following warnings to indicate hazardous situations Symbol Category Mean...
Page 3: ...Safety Guide In official EU languages and others Quickstart Guides Installation and first steps Operation Menu structure Calibration Error messages and recommended actions Test Report 2 2 According to EN 10204 Electronic Documentation www knick international com Manuals software ...
Page 4: ...5 Panel Mounting 16 Electrical Installation 17 Rating Plates Terminal Assignments 18 Wiring of Stratos Pro A201PH 19 Wiring Examples 20 User Interface Keypad 31 Display 32 Signal Colors Display Backlighting 32 Measuring Mode 33 Selecting the Mode Entering Values 34 Display in Measuring Mode 35 Color Coded User Interface 36 Operating Modes 37 Menu Structure of Modes and Functions 38 HOLD Mode 39 Al...
Page 5: ... Measuring Points TAG GROUP 94 Display Backlighting 94 Digital Sensors 96 Memosens Sensors Calibration and Maintenance in the Lab 96 Memosens Sensors Configuring the Device 97 Replacing a Sensor 98 Calibration 100 Selecting a Calibration Mode 101 Zero Adjustment ISFET 102 Automatic Calibration Calimatic 104 Manual Calibration with Buffer Entry 106 Data Entry of Premeasured Sensors 108 Product Cali...
Page 6: ... Line and Accessories 127 Specifications 128 Buffer Tables 134 U1 Specifiable Buffer Set 147 Error Handling 150 Error Messages 151 Decommissioning 154 Disposal 154 Returns 154 Sensoface 155 HART Typical Applications 158 FDA 21 CFR Part 11 159 Electronic Signature Passcodes 159 Audit Trail 159 Glossary 160 Index 165 ...
Page 7: ...s visible damage failure to perform the intended function prolonged storage at temperature of below 30 C 22 F or above 70 C 158 F severe transport stresses Before recommissioning the device a professional routine test must be performed This test should be carried out by the manufacturer at its factory Intended Use Stratos Pro A201 PH is a 2 wire device for pH mV ORP and tempera ture measurement in...
Page 8: ...as the system may behave unexpectedly and put users at risk Control Drawings If installing in hazardous locations observe the information provided on the included Control Drawings Devices Not Intended for Use in Hazardous Locations Devices identified with an N in their product name must not be used in hazardous locations Configuration Replacing components may affect intrinsic safety The modules ar...
Page 9: ...ective hood which is available as accessory provides addi tional protection against direct weather exposure and mechanical damage see page 15 Connection of Sensors Cable Glands For connecting the cables the enclosure provides 3 knockouts for M20x1 5 cable glands 2 knockouts for conduit For quasi stationary installations with Memosens sensors we recom mend using the M12 socket accessory ZU0822 inst...
Page 10: ...indicates the sensor condition see page 155 Data Logger The internal logbook TAN SW A002 can handle up to 100 entries up to 200 with AuditTrail TAN SW A003 see page 119 Two Parameter Sets A B The device provides two parameter sets that can be switched manu ally or via a control input for different process connections or different process conditions For an overview of parameter sets original for co...
Page 11: ... of HOLD mode see page 39 CONTROL Floating digital control input The CONTROL input can be used either for parameter set selection A B or for flow monitoring see page 84 Signal Outputs The device provides two current outputs for transmission of mea sured value and temperature for example Options Additional functions can be enabled by entering a TAN p 123 Introduction 5 6 13 14 10 11 Current input H...
Page 12: ...1 pH mV temp input 3 V 3 V for ISFET adapter 10 11 Output 2 1 2 3 4 RS 485 Control input Current input HOLD input meas el ref el SG RTD RTD ISM GND ISM data ISFET ISFET 3 V 3 V Shield Output 1 2 HART Output 1 HART Output 2 RS 485 A RS 485 B GND Shield Input Input CONTROL CONTROL HOLD HOLD ...
Page 13: ...port Documentation see page 3 Fig Assembling the enclosure 1 Insertable jumper 3x 2 Plate 1x for conduit mount ing Plate between housing and nut 3 Cable tie 3x 4 Hinge pin 1x insertable from either side 5 Enclosure screw 4x 6 Blanking plug 2x non Ex only 7 Reduction sealing insert 1x 8 Cable gland 3x 9 Blanking cap 2x 10 Hex nut 5x 11 Plastic sealing plug 2x for sealing in case of wall mounting ...
Page 14: ...n All dimensions in mm 1 Cable gland 3 x 2 Knockouts for cable gland or conduit 21 5 mm dia 2 knockouts Conduit couplings not included 3 Knockout for pipe mounting 4 x 4 Knockout for wall mounting 2 x 148 148 42 42 42 14 117 41 21 43 1 2 34 80 74 6 2 4 3 ...
Page 15: ...e mounting accessory ZU 0737 All dimensions in mm Fig Pipe mount kit accessory ZU 0274 All dimensions in mm 147 199 91 185 1 2 3 4 1 Hose clamp with worm gear drive to DIN 3017 2 x 2 Pipe mount plate 1 x 3 For vertical or horizontal posts or pipes 4 Self tapping screw 4 x ø40 ø60 ...
Page 16: ...ng Fig Panel mount kit accessory ZU 0738 All dimensions in mm 1 Circumferential sealing 1 x 2 Screws 4 x 3 Position of control panel 4 Span piece 4 x 5 Threaded sleeve 4 x Cutout 138 x 138 mm DIN 43700 30 76 31 1 22 1 2 3 4 5 ...
Page 17: ... Tensile strain Not permitted Only suitable for fixed installation CAUTION Risk of losing the specified ingress protection Fasten the cable glands and screw together the housing correctly Observe the permissible cable diameters and tightening torques Only use original accessories and spare parts NOTICE Strip the insulation from the wires using a suitable tool to prevent damage For stripping length...
Page 18: ...18 Electrical Installation Rating Plates Terminal Assignments Fig Stratos Pro A201N rating plate at outside bottom of front illustrative example Fig Terminal assignments of Stratos Pro A201 ...
Page 19: ...atos Pro A201PH Terminal row 2 Sensor connection MK PH module Fig MK PH module terminal assignments Note When a Memosens sensor is to be con nected to the RS 485 interface terminals 1 4 you must remove the MK PH module Areas for placing the screwdriver to pull out the terminals Fig Terminals device opened back of front unit ...
Page 20: ...20 Wiring Examples Example 1 Measuring task pH temperature glass impedance Sensors example SE 555X 1 NS8N Cable example ZU0318 Jumper Core Shield Sensor s Cable Device ...
Page 21: ...ss impedance ref impedance Sensors example SE555X 1 NS8N equipotential bonding ZU0073 Temperature e g Pt1000 Cable example 2x ZU 0318 ISM GND B C D E F G H A I K meas el ref el SG 3 V 3 V ISM Data RTD GND RTD shield Core Shield Core Shield Sensor s Cable Device ...
Page 22: ...pH temp glass impedance Sensors example SE554X 1 NVPN Cable example CA VP6ST 003A B C D E F G H A I K meas el ref el SG 3 V 3 V ISM GND ISM Data RTD GND RTD shield Green White Jumper Core Shield Shield Clear Red Yellow Green Sensor s Cable Device ...
Page 23: ...ND ISM Data RTD GND RTD shield Example 4 Measuring task pH ORP temp glass impedance ref impedance Sensors example SE555X 1 NVPN equipotential bonding ZU0073 Cable example CA VP6ST 003A Green White Core Shield Shield Clear Red Yellow Green Core Shield Sensor s Cable Device ...
Page 24: ...GND ISM Data RTD GND RTD shield Example 5 Measuring task pH ORP temp glass impedance ref impedance Sensors example PL PETR 120VP pH ORP combo sensor SI Analytics Cable example CA VP6ST 003A Green White Core Shield Shield Clear Red Yellow Green Blue Sensor s Cable Device ...
Page 25: ... V 3 V ISM GND ISM Data RTD GND RTD shield Example 6 Measuring task pH temp safe areas only Sensors example InPro 3300 ISFET Mettler Toledo Cable example ZU 0582 Knick Pink Brown Green White Yellow Green Gray Yellow Jumper Sensor s Cable with amplifier Device ...
Page 26: ...7 Note Do not connect an additional analog sensor Measuring task pH ORP temp glass impedance ref impedance Sensors example ISM digital InPro 4260i Mettler Toledo Cable example AK9 Mettler Toledo Core Shield Sensor s Cable Device ...
Page 27: ...iring Examples Example 8 Note Switch off Sensocheck Measuring task ORP temp glass impedance ref impedance Sensors example SE 564X 1 NS8N Cable example ZU0318 Jumper Core Shield Shield Sensor s Cable Device ...
Page 28: ...g task pH ORP temp glass impedance ref impedance Sensors example SE 533 1 ADIN Cable example CA 003 NAADIN11 Note Connection to RS 485 interface Remove the measuring module Indu Con Yellow White Clear Shield blue Brown Green Sensor s Cable Device ...
Page 29: ...tial Models 18 40 with equipoten tial bonding Models 03 04 with equipoten tial bonding Models 03 04 without equi potential bonding A meas Coax core Coax WH Coax WH Coax WH B ref Coax shield Coax BN Coax BN Coax BN C SG Blue Blue Blue Jumper B C H RTD GND Green Brown Brown Brown I RTD White Green Black Green Black Green Black K Shield Green Yellow Gray Orange Violet Orange Violet Orange Violet Pfau...
Page 30: ...e 11 Memosens Measuring task pH ORP temp glass impedance ref impedance Sensors example Memosens Cable example Memosens cable CA MS 003NAA L Note Connection to RS 485 interface Remove the measuring module Yellow White Shield clear Brown Green Sensor s Cable Device ...
Page 31: ...tion Show error messages Configuration Confirm entries next configuration step Calibration Continue program flow Measuring mode Call menu Arrow keys up down Menu Increase decrease a numeral Menu Selection Arrow keys left right Previous next menu group Number entry Move between digits 1 Display 2 Keypad 3 Nameplate enclosure bottom 2 3 1 MEMO SENS ...
Page 32: ...ay 16 Secondary display 17 Proceed using enter 18 ISM sensor 19 Diagnostics 20 Configuration mode 21 Calibration mode 22 Measuring mode 23 Sensoface 24 Unit symbols 1 2 3 4 5 6 7 8 9 10 11 17 18 19 20 21 22 12 13 14 15 16 23 24 Signal Colors Display Backlighting Red Red blinking Alarm in case of fault display values blink Input error illegal value or wrong passcode Orange HOLD mode Calibration Con...
Page 33: ... default setting Measured value and selection of parameter set A B or flow Measured value and tag number TAG Time and date Note By pressing the meas key in measuring mode you can view the displays for approx 60 sec The device must be configured for the respective measurement task see page 42 Sensoface indicator sensor status Time or flow Mode indicator measuring Hold meas key depressed for calling...
Page 34: ... arrow key 7 Confirm entry by pressing enter To select the operating mode 1 Hold meas key depressed 2 s directly to measuring mode 2 Press menu key the selection menu appears 3 Select operating mode using left right arrow key 4 Press enter to confirm the selected mode Selection menu Selected mode blinks 1 4 3 2 5 7 6 ...
Page 35: ...ain display Press enter to select a display as MAIN DISPLAY the secondary display shows MAIN DISPLAY NO Use the UP DOWN arrow keys to select MAIN DISPLAY YES and confirm by pressing enter The display color changes to white This display is now shown in measuring mode The MAIN DISPLAY is the display which is shown in measuring mode To call the measuring mode from any other mode hold the meas key dep...
Page 36: ...he diagnostic menu appears on turquoise The orange HOLD mode e g during calibration is quickly visible as is the magenta screen which indicates asset management messages for predictive diagnostics such as maintenance request pre alarm and sensor wear The alarm status has a particularly noticeable red display color and is also signaled by flashing display values Invalid inputs or false pass codes c...
Page 37: ... deviation the device can be adjusted In that case the device displays the actual value and internally corrects the measurement error of the sensor Calibration must be repeated at regular intervals The time between the calibration cycles depends on the load on the sensor During calibration the device is in HOLD mode During calibration the device remains in the HOLD mode until it is stopped by the ...
Page 38: ...tment for ISFET only Adjustment of temperature probe Configuring parameter set A Configuring parameter set B Display of measured values for validation simulators Reset TTM for ISM increment autoclaving counter Current source output 1 Current source output 2 Specifying access codes for operating modes Reset to factory setting Enabling an option via TAN Access via code factory setting 5555 Pressing ...
Page 39: ...he installation is ready for operation e g sensor reinstalled located in process HOLD mode display icon HOLD active 21 4 HOLD active Output current mA Output signal for HOLD FIX setting 21 0 mA Output signal for HOLD LAST setting Output signal during HOLD Output signal response LAST The output current is frozen at its last value Recommended for short configuration procedures The process should not...
Page 40: ... alarm be registered and entered in the logbook During an alarm the display blinks the display backlighting turns red Error messages can also be signaled by a 22 mA output current see Configuration 2 sec after the failure event is corrected the alarm status will be deleted External activation of HOLD TAN SW A005 The HOLD mode can be activated from outside by sending a signal to the HOLD input e g ...
Page 41: ...e via the CONTROL input TAN SW A005 min flow max flow The CONTROL input can be used for parameter set selection or for flow measurement pulse principle depending on its assignment in the Configuration menu When preset to flow measurement an alarm can be generated when the measured flow exceeds a specified range specify value default 5 liters h specify value default 25 liters h Power supply 12 24 V...
Page 42: ...ng the parameters Pressing enter opens a menu item Use the arrow keys to edit a value Press enter to confirm save the settings Return to measurement Hold meas key depressed 2 s Select menu group Menu group Code Display Select menu item Sensor selection Menu item 1 Menu item Current output 1 Current output 2 Compensation Display backlighting DSP enter enter enter enter CAUTION Incorrect parameter s...
Page 43: ...ensor selection Current output 1 Current output 1 Current output 2 Current output 2 Compensation Compensation Control input Alarm mode Alarm mode Parameter set selection Setting the clock TAG of measuring point TAG of measuring point GROUP of measuring points GROUP of measuring points Display backlighting ...
Page 44: ...parameter sets Press meas PARSET blinks in the lower line Select parameter set using and keys Press enter to confirm Cancel by pressing meas Configuration External switchover of parameter sets A B TAN SW A005 You can switch between parameter sets A and B by applying a signal to the CONTROL input parameter setting CONTR IN PARSET Parameter set A active 0 2 V AC DC Parameter set B active 10 30 V AC ...
Page 45: ... info displays nominal buffer values manufacturer U1 For specifiable buffer set see Appendix Buffer Tables Enter values for buffer 1 Enter values for buffer 2 omitted for ISM The setting TEMP MEAS AUTO MAN EXT applies to all outputs OUT1 OUT2 display Sensors with deviating zero slope require the Pfaudler option TAN Settings with Sensor STANDARD selected not required for Memosens Pfaudler sensor on...
Page 46: ...SM CIP COUNT ON OFF OFF ON CIP CYCLES 0 9999 CYC 0025 CYC SIP COUNT ON OFF OFF ON SIP CYCLES 0 9999 CYC 0025 CYC AUTOCLAVE ON OFF OFF ON AC CYCLES 0 9999 CYC 0000 CYC CHECK TAG GROUP ON OFF OFF Output 1 OUT1 BEGIN 4 mA END 20 mA BEGIN 4 mA END 20 mA BEGIN 4 mA Select C F at Sensor END 20 mA BEGIN 4 mA END 20 mA FILTERTIME 0 120 SEC 0000 SEC 22mA FAIL ON OFF OFF 22mA FACE ON OFF OFF HOLD MODE LAST ...
Page 47: ...O 0 100 C in 5 C steps with TAN option SW A005 and SENSOR TEMP EXT selected Configuration Select Default Output 2 OUT2 BEGIN 4 mA END 20 mA BEGIN 4 mA END 20 mA BEGIN 4 mA Select C F at Sensor END 20 mA BEGIN 4 mA END 20 mA FILTERTIME 0 120 SEC 0000 SEC 22 mA FAIL ON OFF OFF 22 mA FACE ON OFF OFF HOLD MODE LAST FIX LAST FIX HOLD FIX 04 00 22 00 mA 021 0 mA ...
Page 48: ...T On Off On Hysteresis fixed at 5 of threshold value Control input CNTR_IN IN CONTROL Parameter set switchover PARSET or flow measurement FLOW FLOW FLOW ADJUST 0 20000 pulses liter 12000 pulses liter Alarm ALARM ALA DELAYTIME 0 600 SEC 0010 SEC SENSOCHECK ON OFF OFF FLOW CNTR ON OFF OFF ON FLOW MIN 0 99 9 L h 005 0 L h FLOW MAX 0 99 9 L h 025 0 L h Parameter set selection PARSET PAR Select fixed p...
Page 49: ...sing suitable buffer solutions When you use a Memosens Pfaudler sensor the data will be read from the sensor or will be set to standard values Here you do not have to make entries The respective menu items will be suppressed The nominal ZERO SLOPE values are required for the proper functioning of the sensor monitoring and calibration functions Sensoface Calimatic they do not replace an adjustment ...
Page 50: ...NS Calibration cycle SNS ISM adaptive cal timer ACT SNS ISM adaptive maintenance timer TTM SNS CIP counter SNS SIP counter SNS Autoclaving counter SNS CHECK TAG SNS CHECK GROUP OT1 Process variable OT1 Current start OT1 Current end OT1 Filter time OT1 FAIL 22 mA error messages OT1 FACE 22 mA Sensoface messages OT1 HOLD mode OT1 HOLD FIX current Configuration Template for Copy These parameters cann...
Page 51: ... Temperature coefficient COR Ext temp input COR Current range COR Current start COR Current end IN Parameter set A B or flow IN Flow meter Adjusting pulses liter ALA Delay ALA Sensocheck on off ALA Flow control ALA Minimum flow ALA Maximum flow PAR Parameter set selection CLK Time format TAG Measuring point tag number GROUP Group of measuring points DISPLAY Display backlighting These parameters ca...
Page 52: ... indicated by the SNS code Press enter to select menu edit using arrow keys see next page Confirm and proceed by pressing enter 6 Exit Press meas key until the meas mode indicator is displayed Select sensor type Select type of temp probe Temperature unit Temp detection during measurement Manual temperature Temp detection during calibration Manual temperature Calibration mode AUTO Buffer set Calibr...
Page 53: ...enter to confirm 1000 PT 100 PT Balco 3 kΩ 8 55 NTC 30 NTC Temperature unit Select C or F using keys Press enter to confirm Temp detection during measurement Select mode using keys AUTO Measured by sensor MAN Direct input of temperature no measure ment see next step EXT Temperature speci fied via current input only if TAN E enabled Press enter to confirm Manual temperature Modify digit using keys ...
Page 54: ... SNS code Press enter to select menu edit using arrow keys see next page Confirm and proceed by pressing enter 6 Exit Press meas key until the meas mode indicator is displayed Select sensor type Select type of temp probe Temperature unit Temp detection during measurement Manual temperature Temp detection during calibration Manual temperature Calibration mode AUTO Buffer set Calibration timer Calib...
Page 55: ...ify digit using keys select next digit using keys Press enter to confirm Calibration mode Select CALMODE using keys AUTO Calibration with Calimatic buffer set recognition MAN Manual entry of buffer solutions DAT Input of adjustment data of premeasured sensors Press enter to confirm AUTO Buffer set Select buffer set using keys see buf fer tables for nominal values Press enter to confirm 01 13 U1 se...
Page 56: ... enter to select menu edit using arrow keys see next page Confirm and proceed by pressing enter 6 Exit Press meas key until the meas mode indicator is displayed Select sensor type Select type of temp probe Temperature unit Temp detection during measurement Manual temperature Temp detection during calibration Manual temperature Calibration mode AUTO Buffer set Calibration timer Calibration cycle CI...
Page 57: ...sensor wear Calibration cycle Only with FIX ADAPT Modify digit using keys select next digit using keys Press enter to confirm 5 Note for the calibration timer When Sensocheck has been activated in the Configuration Alarm menu the expiration of the calibration interval is indicated by Sensoface Display Status Over 80 of the calibration interval has already past The calibration interval has been exc...
Page 58: ...aptive maintenance timer CIP SIP cycles Autoclaving counter 1 Press menu key 2 Select CONF using keys press enter 3 Select parameter set using keys press enter 4 Select SENSOR menu using keys press enter 5 All items of this menu group are indicated by the SNS code Press enter to select menu edit using arrow keys see next page Confirm and proceed by pressing enter 6 Exit Press meas key until the me...
Page 59: ...ither read automatically from the sensor settings or can be specified manually max 9999 days Stressing influences temperature measurement in extreme ranges shorten the timer interval The adaptive cal timer is reset after each calibration You can configure the current outputs so that a Sensoface message generates a 22 mA error signal see page 75 Menu item Action Choices Adaptive cal timer ACT Selec...
Page 60: ...M Adaptive maintenance timer CIP SIP cycles Autoclaving counter 1 Press menu key 2 Select CONF using keys press enter 3 Select parameter set using keys press enter 4 Select SENSOR menu using keys press enter 5 All items of this menu group are indicated by the SNS code Press enter to select menu edit using arrow keys see next page Confirm and proceed by pressing enter 6 Exit Press meas key until th...
Page 61: ... specified manually max 9999 days Stressing influences temperature measurement in extreme ranges shorten the timer interval You can configure the current outputs so that a Sensoface message generates a 22 mA error signal see page 75 Menu item Action Choices Adaptive mainte nance timer TTM Select using keys AUTO The interval stored in the ISM sensor is used default MAN The interval is specified man...
Page 62: ...y the SNS code Press enter to select menu edit using arrow keys see next page Confirm and proceed by pressing enter 6 Exit Press meas key until the meas mode indicator is displayed Select sensor type Select type of temp probe Temperature unit Temp detection during measurement Temp detection during calibration Calibration mode AUTO Buffer set Calibration timer Calibration cycle CIP cleaning cycles ...
Page 63: ... or OFF using keys When switched on the cycles will be entered in the extended logbook TAN SW A003 but will not be counted Press enter to confirm Logging the cleaning and sterilization cycles with connected sensor helps measuring the load on the sensor Suitable for biochemical applications process temp approx 0 50 C 32 122 F CIP temp 55 C 131 F SIP temp 115 C 239 F Note A CIP or SIP cycle is only ...
Page 64: ...ning cycles Sterilization cycle counter Sterilization cycles Autoclaving counter CHECK TAG CHECK GROUP 1 Press menu key 2 Select CONF using keys press enter 3 Select parameter set using keys press enter 4 Select SENSOR menu using keys press enter 5 All items of this menu group are indicated by the SNS code Press enter to select menu edit using arrow keys see next page Confirm and proceed by pressi...
Page 65: ...enter to confirm 0 9999 CYC 0000 CYC Sterilization cycle counter Select ON or OFF using keys Press enter to confirm ON OFF Sterilization cycles Only with CIP COUNT ON Enter value using keys Press enter to confirm 0 9999 CYC 0000 CYC Logging the cleaning and sterilization cycles with connected sensor helps measuring the load on the sensor Suitable for biochemical applications process temp approx 0 ...
Page 66: ...libration timer TTM Adaptive maintenance timer CIP SIP cycles Autoclaving counter CHECK TAG CHECK GROUP 1 Press menu key 2 Select CONF using keys press enter 3 Select parameter set using keys press enter 4 Select SENSOR menu using keys press enter 5 All items of this menu group are indicated by the SNS code Press enter to select menu edit using arrow keys see next page Confirm and proceed by press...
Page 67: ...ansmitter The transmitter displays INCREMENT AUTOCLAVE CYCLE as confir mation You can configure the current outputs so that a Sensoface message generates a 22 mA error signal see page 75 Menu item Action Choices Autoclaving counter Select using keys ON The cycles are specified manually 0 9999 Press enter to confirm With the autoclaving counter switched on you must increment the count after each au...
Page 68: ...NS code Press enter to select menu edit using arrow keys see next page Confirm and proceed by pressing enter 6 Exit Press meas key until the meas mode indicator is displayed 1 2 4 enter 6 enter enter meas 3 MEMO SENS 5 Select sensor type Temperature unit Temp detection during measurement Temp detection during calibration Calibration mode AUTO Buffer set Calibration timer Calibration cycle CIP SIP ...
Page 69: ...or group of measuring points GROUP in the sensor TAG and GROUP can be specified by the calibration tool or automatically entered by the transmitter When connecting an MS sensor to the transmitter it can be checked if the sensor contains the correct TAG or belongs to the correct GROUP If not a message will be generated Sensoface gets sad and the display backlight ing turns purple magenta The sad Se...
Page 70: ...de Press enter to select menu edit using arrow keys see next page Confirm and proceed by pressing enter 6 Exit Press meas key until the meas mode indicator is displayed 6 5 Process variable Current start Current end Time averaging filter Output current for error message Output current for Sensoface Output current during HOLD Output current for HOLD FIX Current Output 1 Output current range Current...
Page 71: ...P Current start Modify digit using keys select next digit using keys Press enter to confirm 0 42 5 rH Current end Enter value using keys Press enter to confirm 0 42 5 rH Assignment of measured values Current start and current end Output current pH 20 4 14 0 20 4 7 5 mA Example 1 Range pH 0 14 mA Output current 7 5 5 Example 2 Range pH 5 7 Advantage Higher resolution in range of interest pH ...
Page 72: ...r 5 All items of this menu group are indicated by the OT1 code Press enter to select menu edit using arrow keys see next page Confirm and proceed by pressing enter 6 Exit Press meas key until the meas mode indicator is displayed 6 5 Process variable Current start Current end Time averaging filter Output current during error message Output current for Sensoface Output current during HOLD Output cur...
Page 73: ...jump at the input 100 the output level is at 63 after the time interval has been reached The time interval can be set from 0 to 120 sec If the time interval is set to 0 sec the current output directly follows the input Note The filter only acts on the current output not on the display or the limit value During HOLD the filter is not applied This prevents a jump at the output Time interval 0 120 s ...
Page 74: ...or Sensoface Output current during HOLD Output current for HOLD FIX 1 Press menu key 2 Select CONF using keys press enter 3 Select parameter set using keys press enter 4 Select OUT1 menu using keys press enter 5 All items of this menu group are indicated by the OT1 code Press enter to select menu edit using arrow keys see next page Confirm and proceed by pressing enter 6 Exit Press meas key until ...
Page 75: ...r Sensoface FACE In the case of a Sensoface alert FACE the current output is set to 22 mA Select ON or OFF using keys Press enter to confirm pH Temp Display Error 22 mA Sensoface alert 22 mA Error messages and Sensoface alerts can be set separately for both current outputs This allows for example signaling error messages only over current output 1 and Sensoface alerts only over current output 2 4 ...
Page 76: ...s menu key 2 Select CONF using keys press enter 3 Select parameter set using keys press enter 4 Select OUT1 menu using keys press enter 5 All items of this menu group are indicated by the OT1 code Press enter to select menu edit using arrow keys see next page Confirm and proceed by pressing enter 6 Exit Press meas key until the meas mode indicator is displayed Current Output 1 Output current durin...
Page 77: ... at the output Select using Press enter to confirm LAST FIX Output current for HOLD FIX Only with FIX selected Enter current which is to flow at the output during HOLD Enter value using keys Press enter to confirm 04 00 22 00 mA 21 00 mA HOLD active 21 4 HOLD active Output current mA Output signal for HOLD FIX setting 21 0 mA Output signal for HOLD LAST setting Output signal during HOLD 5 ...
Page 78: ...er to select menu edit using arrow keys see next page Confirm and proceed by pressing enter 6 Exit Press meas key until the meas mode indicator is displayed 6 5 Configuration Current Output 2 Process variable Current start Current end Process variable Current start Current end Time averaging filter Output current for error message Output current for Sensoface Output current during HOLD Output curr...
Page 79: ...Choices Process variable Select using keys PH pH value ORP Redox potential RH rH value with pH ORP sensor TMP Temperature Press enter to confirm PH ORP RH TMP All the following adjustments are made as for current output 1 see page 70 5 ...
Page 80: ...ted by the COR code Press enter to select menu edit using arrow keys see next page Confirm and proceed using enter 6 Exit Press meas key until the meas mode indicator is displayed 6 5 Configuration Temperature compensation for process medium Current input external temp measurement Current range Current start Current end Temperature Compensation TC process medium Linear ultrapure water table ...
Page 81: ...alue using keys Press enter to confirm Temp compensation table Confirm safety prompt then enter values 5 C step size When you have selected temperature compensa tion via table USERTAB you can enter values for a TC table from 0 to 95 C in 5 K steps The analyzer displays temperature values in 5 C steps You must spec ify the percent deviation of the measured value from each of these tem perature valu...
Page 82: ...u group are indicated by the COR code Press enter to select menu edit using arrow keys see next page Confirm and proceed using enter 6 Exit Press meas key until the meas mode indicator is displayed 6 5 Temperature compensation for process medium Current input external temp measurement Current range Current start Current end Temperature Compensation Current input external temp measurement ...
Page 83: ...uring con figuration SENSOR Select ON or OFF using keys Press enter to confirm Current range Select desired range using keys Press enter to confirm Current start Modify digit using keys select next digit using keys Press enter to confirm Input range Current end Enter value using keys Press enter to confirm Input range 5 ...
Page 84: ...3 Select parameter set using press enter 4 Select CNTR_IN menu using keys press enter 5 All items of this menu group are indicated by the IN code Press enter to select menu edit using arrow keys see next page Confirm and proceed using enter 6 Exit Press meas key until the meas mode indicator is displayed CONTROL input function PARSET FLOW 6 ...
Page 85: ...CONTROL input External switchover of parameter sets The parameter set A B can be activated from outside by sending a signal to the CONTROL input e g from the process control system Configuration 5 Parameter set A active 0 2 V AC DC Parameter set B active 10 30 V AC DC e g process control system Max 30 V AC DC CONTROL input 13 14 Stratos Pro A2 ...
Page 86: ... CONF using keys press enter 3 Select parameter set using press enter 4 Select CNTR_IN menu using keys press enter 5 All items of this menu group are indicated by the IN code Press enter to select menu edit using arrow keys see next page Confirm and proceed using enter 6 Exit Press meas key until the meas mode indicator is displayed ...
Page 87: ...erated Configuration Menu item Action Choices Select function of CONTROL input Select using keys Press enter to confirm PARSET selecting parameter set A B via signal at CONTROL input Flow for connecting a pulse output flow meter Adjust to flow meter With Flow selected you must adjust the device to the flow meter used Enter value using arrow keys Press enter to confirm 12000 pulses liter Display Fl...
Page 88: ... ALARM menu using keys press enter 5 All items of this menu group are indicated by the ALA code Press enter to select menu edit using arrow keys see next page Confirm and proceed using enter 6 Exit Press meas key until the meas mode indicator is displayed 6 5 Delay Sensocheck CONTROL input For flow monitoring Max flow alarm For flow monitoring Min flow alarm ...
Page 89: ...hting to red and the 22 mA signal if configured Configuration Menu item Action Choices Alarm delay Enter value using keys Press enter to confirm 0 600 SEC 010 SEC Sensocheck Select Sensocheck continuous monitoring of glass and reference electrode Select ON or OFF using keys Press enter to confirm At the same time Sensoface is activated With OFF Sensoface is also switched off ON OFF 5 ...
Page 90: ...t ALARM menu using keys press enter 5 All items of this menu group are indicated by the ALA code Press enter to select menu edit using arrow keys see next page Confirm and proceed using enter 6 Exit Press meas key until the meas mode indicator is displayed 5 Delay Sensocheck CONTROL input For flow monitoring Max flow alarm For flow monitoring Min flow alarm 6 ...
Page 91: ...te an alarm when assigned to FLOW flow monitoring in the CONF menu FLOW CNTR Flow measurement allows monitoring the minimum and maximum flow pulse counter Alarm Minimum flow FLOW MIN Specify value Default 05 00 liters h Alarm Maximum flow FLOW MAX Specify value Default 25 00 liters h 5 ...
Page 92: ...NF using press enter 3 Select parameter set A using keys press enter 4 Select CLOCK using keys press enter 5 All items of this menu group are indicated by the CLK code Press enter to select menu edit using arrow keys see next page Confirm and proceed by pressing enter 6 Exit Press meas key until the meas mode indicator is displayed ...
Page 93: ...eal time clock In measuring mode the time is shown in the lower display When using digital sensors the calibration data is written in the sensor head In addition the logbook entries cf Diagnostics are provided with a time stamp Note There is no automatic switchover from winter to summer time Be sure to manually adjust the time ...
Page 94: ...Press menu key 2 Select CONF using press enter 3 Select parameter set A using keys press enter 4 Select TAG or DISPLAY using keys press enter 5 All items of this menu group are indicated by the TAG or DSP code Press enter to select menu edit using arrow keys see next page Confirm and proceed by pressing enter 6 Exit Press meas key until the meas mode indicator is displayed ...
Page 95: ... be switched on in the Configuration in two steps as TAG and GROUP if required When no measuring point or group of measuring points is saved in the sensor e g when using a new sensor Stratos enters its own TAG and GROUP When sensor veri fication is switched off Stratos always enters its own measuring point and group A possibly existing TAG GROUP will be overwritten Menu item Action Choices Measuri...
Page 96: ...gital sensors can be calibrated and maintained in the lab This considerably simplifies on site maintenance Memosens Sensors Calibration and Maintenance in the Lab The MemoSuite software allows calibrating Memosens sensors under reproducible conditions at a PC in the lab The sensor param eters are registered in a database Documenting and archiving meet the demands of FDA CFR 21 Part 11 Detailed rep...
Page 97: ...u can display the error text in the bottom line using the keys The comprehensive display options of the software allow drawing conclusions on the aging behavior of the sensors at one glance and making a reliable forecast for predictive maintenance The software is available as Basic calibration function or Advanced with sensor database version www knick de Calibration history of several sensors His...
Page 98: ...S View sensor information using keys press enter to confirm Display color changes to green Sensoface is friendly when the sensor data are okay Go to measuring mode Press meas info or enter After 60 sec the device automatically returns to measuring mode time out Replacing a Sensor A sensor should only be replaced during HOLD mode to prevent un intended reactions of the outputs or contacts When you ...
Page 99: ...ary messages are dis played during the replacement but neither output to the alarm contact nor entered in the log book Wait until the sensor data are displayed MEMO SENS Check sensor data MEMO SENS View sensor informa tion using keys press enter to confirm You can view the sensor manu facturer and type serial number and last calibration date Check measured values then exit HOLD Hit meas key Return...
Page 100: ...only operate properly when the buffer solutions used correspond to the configured set Other buffer solutions even those with the same nominal values may demonstrate a different temperature response This leads to measurement errors When using ISFET sensors or sensors with a zero point other than pH 7 the nominal zero point must be adjusted each time a new sensor is connected This is important if yo...
Page 101: ... of premeasured electrode data CAL_ORP ORP calibration P_CAL Product calibration calibration with sampling ISFET ZERO Zero adjustment Required for ISFET sensors subsequently you can conduct either a one or a two point calibration CAL_RTD Temperature probe adjustment Selecting a Calibration Mode To preset CAL_PH CONF menu configuration 1 Hold meas key depressed 2 s measuring mode 2 Press menu key t...
Page 102: ...k Select Calibration Press enter to proceed Ready for calibration Hourglass blinks Display 3 sec Now the device is in HOLD mode Immerse sensor in a pH 7 00 buffer Enter the temperature corrected pH value in the range 6 50 to 7 50 using the arrow keys see buffer table Confirm with enter If the zero offset of the sensor is too large 200 mV a CAL ERR error message is gener ated In that case the elect...
Page 103: ... be determined with a complete 2 point calibration Use the arrow keys to select Repeat repeat calibration or Measuring Confirm by pressing enter Place sensor in process Press enter to end zero calibration After end of calibra tion the outputs re main in HOLD mode for a short time Note for zero adjustment After having adjusted the zero offset be sure to calibrate the sensor following one of the pro...
Page 104: ...ion method CAL_PH Press enter to proceed Display 3 sec Now the device is in HOLD mode Remove the sensor and temperature probe clean them and im merse them in the first buffer solution in any order Start by pressing enter When manual input of temperature has been configured the temp value in the display blinks and can be edited using the arrow keys Buffer recognition While the hourglass icon is bli...
Page 105: ...essing enter However this reduces calibration accuracy Display for 1 point cal Sensoface is active End by pressing enter 2 point calibration Immerse sensor and temperature probe in the second buffer solution Press enter to start The calibration process runs as for the first buffer Retract sensor and temp probe out of second buffer rinse off re install Press enter to proceed The slope and asym metr...
Page 106: ...lass blinks Display 3 sec Now the device is in HOLD mode Remove the sensor and temperature probe clean them and im merse them in the first buffer solution Press enter to start When manual input of temperature has been configured the temp value in the display blinks and can be edited using the arrow keys Enter the pH value of your buffer solution for the proper temperature While the hourglass icon ...
Page 107: ...ec by pressing enter However this reduces calibration accuracy Display for 1 point cal Sensoface is active End by pressing enter 2 point calibration Immerse sensor and temperature probe in the second buffer solution Enter pH value Press enter to start The calibration process runs as for the first buffer Rinse sensor and tem perature probe and reinstall them Press enter to proceed Display of slope ...
Page 108: ...and in the laboratory Display Action Remark Select Calibration Press enter to proceed Data Input Ready for calibration Hourglass blinks Display 3 sec Now the device is in HOLD mode Enter asymmetry potential mV Press enter to proceed Enter slope The device displays the new slope and asymmetry potential at 25 C Sensoface is active Use the arrow keys to select MEAS exit REPEAT Press enter to proceed ...
Page 109: ...pe mV pH at 25 C mV pH 78 46 2 80 47 4 82 48 5 84 49 7 86 50 9 88 52 1 90 53 3 92 54 5 94 55 6 96 56 8 98 58 0 100 59 2 102 60 4 Converting asymmetry potential to sensor zero point Sensor zero Asymmetry potential Slope ZERO 7 ZERO VAS S VAS mV S mV pH ...
Page 110: ...turns to measuring mode The calibration mode indicator blinks 2 In the second step you enter the measured sample value in the device From the difference between the stored measured value and entered sample value the device calculates the new asymmetry potential If the sample is invalid you can take over the value stored during sampling In that case the old calibration values are stored Afterwards ...
Page 111: ...in HOLD mode The stored value is displayed blinking and can be overwritten with the measured sample value Press enter to proceed Display of new asym metry potential based on 25 C Sensoface is active To exit calibration Select MEAS then press enter To repeat calibra tion Select REPEAT then press enter End of calibration After end of calibration the outputs remain in HOLD mode for a short time Produ...
Page 112: ... hydrogen electrode In that case the temperature corrected potential see table of the reference electrode used must be entered during calibration During measurement this value is then added to the ORP measured Please make sure that measurement and calibration temperature are the same since the temperature behavior of the reference electrode is not automatically taken into account 0 10 20 25 30 40 ...
Page 113: ...x buffer Display 3 sec Now the device is in HOLD mode Enter setpoint value for redox buffer Press enter to proceed The ORP delta value is displayed based on 25 C Sensoface is active Press enter to proceed To repeat calibration Select REPEAT To exit calibration Select MEAS then press enter After end of calibra tion the outputs re main in HOLD mode for a short time ...
Page 114: ...he measured temperature value Maximum difference 10 K Press enter to proceed Display of actual temperature un compensated in the lower display The corrected tempera ture value is displayed Sensoface is active To exit calibration Select MEAS then press enter To repeat calibration Select REPEAT then press enter After calibration is ended the device will switch to measuring mode After end of calibra ...
Page 115: ...e pH ORP mV or temperature The meas mode indicator lights and the active parameter set A B is indicated By pressing the meas key you can step through the following dis plays When no key has been pressed for 60 sec the device returns to the MAIN DISPLAY 1 Selecting the parameter set if set to manual in the configuration Select the desired parameter set using the arrow keys PARSET A or PARSET B blin...
Page 116: ...ying device type software version serial number Access to diagnostics can be protected with a passcode SERVICE menu Note HOLD is not active during Diagnostics mode Action Key Remark Activate Diagnostics Press menu key to call the selection menu Display color changes to turquoise Select DIAG using keys confirm by pressing enter Select diagnostics option Use keys to select from CALDATA SENSOR SELFTE...
Page 117: ...LOPE NEXT_CAL The selected parameter is shown in the main display Press meas to return to measurement Display of sensor data For analog sensors the type is displayed STANDARD ISFET Not applicable for digital transmitters MSPH For digital sensors the manufacturer type serial number and last calibration date is displayed In each case Sensoface is active Display data using keys return by pressing ent...
Page 118: ...s enter to proceed 2 RAM test Hourglass blinks then display of PASS or FAIL Press enter to proceed 3 EEPROM test Hourglass blinks then display of PASS or FAIL Press enter to proceed 4 FLASH test Hourglass blinks then display of PASS or FAIL Press enter to proceed 5 Module test Hourglass blinks then display of PASS or FAIL Press enter or meas to return to measuring mode ...
Page 119: ...s to view the corresponding message text If the display is set to the message text you can search for a particular message using the keys Press to display the date and time Press meas to return to measurement Extended logbook Audit Trail TAN SW A003 Using the keys you can scroll backwards and forwards through the extended logbook entries 000 199 000 being the last entry Display CFR Audit Trail als...
Page 120: ...ntenance timer DLI Dynamic Life Time Indicator The selected parameter is shown in the main display Press meas to return to measurement Display mV_pH for validation sensor can be immersed in a calibra tion solution for example or the device is checked by using a simulator Display of remaining dynamic lifetime only for digital sensors however not for MEMOSENS Display of sensor operating time for dig...
Page 121: ... DEFAULT Resetting the device to factory settings OPTION Enabling options via TAN Note HOLD is active during Service mode Action Key Display Remark Activate Service Press menu key to call the selection menu Select SERVICE using keys press enter to confirm Passcode Enter passcode 5555 for service mode using the keys Press enter to confirm Display In service mode the following icons are displayed di...
Page 122: ...ting the adaptive maintenance timer Here the interval is reset to its initial value To do so select TTM RESET YES and confirm by pressing enter Incrementing the autoclaving counter After having completed an autoclaving process you must increment the autoclaving count To do so select YES and confirm by pressing enter The device confirms with INCREMENT AUTOCLAVE CYCLE Specifying the current at outpu...
Page 123: ... the Service passcode to 5555 Reset to factory settings In the SERVICE DEFAULT menu you can reset the device to factory settings NOTICE After a reset to factory setting the device must be reconfigured completely including the sensor parameters Option request Communicate the serial number and hardware soft ware version of your device to the manufacturer These data can be viewed in the Diagnostics V...
Page 124: ...ng States Operating status OUT 1 OUT 2 Time out Measuring DIAG 60 s CAL No CONF 20 min SERVICE 20 min SERVICE OUT 1 20 min SERVICE OUT 2 20 min HOLD No Explanation as configured Last Fix or Last Off active manual ...
Page 125: ... activate the function check HOLD mode on the device as follows Opening the Calibration menu Opening the Service menu Opening the Confiuration menu Repair The Stratos Pro and the measuring modules cannot be repaired by the user To request a repair please contact Knick Elektronische Messgeräte GmbH Co KG by visiting www knick de Maintenance and Repair ...
Page 126: ...pt 336 Repeater power supply Ex 24 V AC DC HART output 4 20 mA WG 21 A7 Opt 336 470 Stratos Pro A201B Zone 2 Repeater power supply non Ex 24 V DC output 4 20 mA IsoAmp PWR B 10116 Repeater power supply non Ex 24 V DC HART output 0 4 20 mA 0 10 V IsoAmp PWR A 20100 A201B X Supply Units and Connection Connection to Supply Units Supply unit 1 Supply unit 2 Hazardous area requirements for cable Betwee...
Page 127: ... Dual COND 2x2 electrode sensors analog N CC pH ORP value ISM digital per TAN Measuring module PH F G Cond 2 4 electrode Measuring module COND Conductivity electrodeless Measuring module CONDI Oxygen ISM digital and traces per TAN Measuring module OXY D F Options Without 2nd current output 0 With 2nd current output 1 TAN options HART SW A001 A Logbook SW A002 B Extended logbook Audit Trail SW A003...
Page 128: ...value 0 02 TC 0 002 pH K mV value 1 mV TC 0 1 mV K pH sensor standardization pH calibration Operating modes BUF Calibration with automatic buffer recognition Calimatic MAN Manual calibration with entry of individual buffer values DAT Data entry of premeasured electrodes Product calibration Calimatic buffer sets 01 Mettler Toledo 2 00 4 01 7 00 9 21 02 Knick CaliMat 2 00 4 00 7 00 9 00 12 00 03 Cib...
Page 129: ...uation of zero slope response time calibration interval wear Sensocheck can be disabled Temperature input Pt100 Pt1000 NTC 30 kΩ NTC 8 55 kΩ Balco 3 kΩ 2 wire connection adjustable Measuring range Pt 100 Pt 1000 20 0 200 0 C 4 392 F NTC 30 kΩ 20 0 150 0 C 4 302 F NTC 8 55 kΩ Mitsubishi 10 0 130 0 C 14 266 F Balco 3 kΩ 20 0 130 0 C 4 266 F Adjustment range 10 K Resolution 0 1 C 0 1 F Measurement er...
Page 130: ... Current input 0 4 20 mA 50 Ω for external temperature signal Start end of scale Configurable 20 200 C 4 392 F Characteristic Linear Resolution approx 0 05 mA Measurement error 1 3 1 of current value 0 1 mA HOLD input TAN Galvanically isolated optocoupler Function Switches device to HOLD mode Switching voltage 0 2 V AC DC HOLD inactive 10 30 V AC DC HOLD active CONTROL input TAN Galvanically isola...
Page 131: ...egment with icons Main display Character height approx 22 mm unit symbols approx 14 mm Secondary display Character height approx 10 mm Text line 14 characters 14 segments Sensoface 3 status indicators friendly neutral sad face Status indicators meas cal conf diag Further icons for configuration and messages Alarm indication Display blinks and red backlighting Keypad Keys meas menu info 4 cursor ke...
Page 132: ...ptionally available add on functions Data retention Parameters calibration data and logbook 10 years EEPROM Housing Molded enclosure glass fiber reinforced Front unit material PBT Rear unit material PC Mounting Wall pipe post or panel mounting Color Gray RAL 7001 Ingress protection IP66 IP67 TYPE 4X outdoor with pressure compensation when the device is closed Flammability UL 94 V 0 for external pa...
Page 133: ...Relative humidity 5 95 Supply voltage 14 30 V Transport and storage Transport storage temperature 30 70 C 22 158 F EMC Emitted interference Class A industrial applications 4 Immunity to interference Industrial applications User defined 1 At rated operating conditions 2 1 digit 3 Plus sensor error 4 This equipment is not designed for domestic use and is unable to guarantee adequate protection of th...
Page 134: ...32 20 2 00 4 00 7 02 9 26 25 2 00 4 01 7 00 9 21 30 1 99 4 01 6 99 9 16 35 1 99 4 02 6 98 9 11 40 1 98 4 03 6 97 9 06 45 1 98 4 04 6 97 9 03 50 1 98 4 06 6 97 8 99 55 1 98 4 08 6 98 8 96 60 1 98 4 10 6 98 8 93 65 1 99 4 13 6 99 8 90 70 1 99 4 16 7 00 8 88 75 2 00 4 19 7 02 8 85 80 2 00 4 22 7 04 8 83 85 2 00 4 26 7 06 8 81 90 2 00 4 30 7 09 8 79 95 2 00 4 35 7 12 8 77 Buffer Tables ...
Page 135: ... 12 39 10 2 01 4 02 7 04 9 11 12 26 15 2 00 4 01 7 02 9 05 12 13 20 2 00 4 00 7 00 9 00 12 00 25 2 00 4 01 6 99 8 95 11 87 30 2 00 4 01 6 98 8 91 11 75 35 2 00 4 01 6 96 8 88 11 64 40 2 00 4 01 6 96 8 85 11 53 50 2 00 4 01 6 96 8 79 11 31 60 2 00 4 00 6 96 8 73 11 09 70 2 00 4 00 6 96 8 70 10 88 80 2 00 4 00 6 98 8 66 10 68 90 2 00 4 00 7 00 8 64 10 48 ...
Page 136: ... 9 99 25 2 08 4 02 6 98 9 95 30 2 06 4 00 6 96 9 89 35 2 06 4 01 6 95 9 85 40 2 07 4 02 6 94 9 81 45 2 06 4 03 6 93 9 77 50 2 06 4 04 6 93 9 73 55 2 05 4 05 6 91 9 68 60 2 08 4 10 6 93 9 66 65 2 07 4 10 6 92 9 61 70 2 07 4 11 6 92 9 57 75 2 04 4 13 6 92 9 54 80 2 02 4 15 6 93 9 52 85 2 03 4 17 6 95 9 47 90 2 04 4 20 6 97 9 43 95 2 05 4 22 6 99 9 38 extrapolated Buffer Tables ...
Page 137: ... 01 12 46 30 1 68 4 015 6 985 9 97 12 30 35 1 69 4 025 6 98 9 93 12 13 40 1 69 4 03 6 975 9 89 11 99 45 1 70 4 045 6 975 9 86 11 84 50 1 705 4 06 6 97 9 83 11 71 55 1 715 4 075 6 97 9 83 11 57 60 1 72 4 085 6 97 9 83 11 45 65 1 73 4 10 6 98 9 83 11 45 70 1 74 4 13 6 99 9 83 11 45 75 1 75 4 14 7 01 9 83 11 45 80 1 765 4 16 7 03 9 83 11 45 85 1 78 4 18 7 05 9 83 11 45 90 1 79 4 21 7 08 9 83 11 45 95...
Page 138: ...11 11 705 55 1 715 4 075 6 834 8 985 11 574 60 1 723 4 091 6 836 8 962 11 449 70 1 743 4 126 6 845 8 921 80 1 766 4 164 6 859 8 885 90 1 792 4 205 6 877 8 850 95 1 806 4 227 6 886 8 833 Note The actual pH values of the individual batches of the reference materials are documented in a certificate of an accredited laboratory This certificate is supplied with the respective buffers Only these pH S va...
Page 139: ...17 15 4 00 7 036 10 11 20 4 00 7 016 10 05 25 4 01 7 00 10 00 30 4 01 6 987 9 96 35 4 02 6 977 9 92 40 4 03 6 97 9 88 45 4 05 6 965 9 85 50 4 06 6 964 9 82 55 4 07 6 965 9 79 60 4 09 6 968 9 76 65 4 10 6 98 9 71 70 4 12 7 00 9 66 75 4 14 7 02 9 63 80 4 16 7 04 9 59 85 4 18 7 06 9 56 90 4 21 7 09 9 52 95 4 24 7 12 9 48 ...
Page 140: ... 10 26 20 2 00 4 00 7 02 10 13 25 2 00 4 01 7 00 10 00 30 1 99 4 01 6 99 9 87 35 1 99 4 02 6 98 9 74 40 1 98 4 03 6 97 9 61 45 1 98 4 04 6 97 9 48 50 1 98 4 06 6 97 9 35 55 1 98 4 08 6 98 60 1 98 4 10 6 98 65 1 99 4 13 6 99 70 2 00 4 16 7 00 75 2 00 4 19 7 02 80 2 00 4 22 7 04 85 2 00 4 26 7 06 90 2 00 4 30 7 09 95 2 00 4 35 7 12 Buffer Tables ...
Page 141: ...30 1 99 4 01 6 99 9 97 11 90 35 1 98 4 02 6 98 9 92 11 80 40 1 98 4 03 6 97 9 86 11 70 45 1 97 4 04 6 97 9 83 11 60 50 1 97 4 05 6 97 9 79 11 51 55 1 98 4 06 6 98 9 75 11 42 60 1 98 4 08 6 98 9 72 11 33 65 1 98 4 10 6 99 9 69 11 24 70 1 99 4 12 7 00 9 66 11 15 75 1 99 4 14 7 02 9 63 11 06 80 2 00 4 16 7 04 9 59 10 98 85 2 00 4 18 7 06 9 56 10 90 90 2 00 4 21 7 09 9 52 10 82 95 2 00 4 24 7 12 9 48 ...
Page 142: ... 4 01 6 99 8 95 11 81 35 2 00 4 02 6 98 8 90 11 63 40 2 01 4 03 6 97 8 86 11 47 45 2 01 4 04 6 97 8 83 11 39 50 2 00 4 05 6 96 8 79 11 30 55 2 00 4 07 6 96 8 77 11 13 60 2 00 4 08 6 96 8 74 10 95 65 2 00 4 10 6 99 8 70 10 95 70 2 00 4 12 7 00 8 67 10 95 75 2 00 4 14 7 02 8 64 10 95 80 2 00 4 16 7 04 8 62 10 95 85 2 00 4 18 7 06 8 60 10 95 90 2 00 4 21 7 09 8 58 10 95 95 2 00 4 24 7 12 8 56 10 95 V...
Page 143: ...9 23 12 75 30 1 10 4 65 6 78 9 18 12 61 35 1 10 4 65 6 77 9 13 12 45 40 1 10 4 66 6 76 9 09 12 29 45 1 10 4 67 6 76 9 04 12 09 50 1 11 4 68 6 76 9 00 11 89 55 1 11 4 69 6 76 8 96 11 79 60 1 11 4 70 6 76 8 92 11 69 65 1 11 4 71 6 76 8 90 11 56 70 1 11 4 72 6 76 8 88 11 43 75 1 11 4 73 6 77 8 86 11 31 80 1 12 4 75 6 78 8 85 11 19 85 1 12 4 77 6 79 8 83 11 09 90 1 13 4 79 6 80 8 82 10 99 95 1 13 4 82...
Page 144: ...30 1 99 4 01 6 99 8 95 10 93 35 1 98 4 02 6 98 8 90 10 86 40 1 98 4 03 6 97 8 85 10 80 45 1 97 4 04 6 97 8 82 10 73 50 1 97 4 05 6 97 8 78 10 67 55 1 98 4 06 6 98 8 75 10 61 60 1 98 4 08 6 98 8 72 10 55 65 1 98 4 10 6 99 8 70 10 49 70 1 99 4 12 7 00 8 67 10 43 75 1 99 4 14 7 02 8 64 10 38 80 2 00 4 16 7 04 8 62 10 33 85 2 00 4 18 7 06 8 60 10 28 90 2 00 4 21 7 09 8 58 10 23 95 2 00 4 24 7 12 8 56 ...
Page 145: ...30 1 99 4 01 6 00 8 95 10 93 35 1 98 4 02 6 00 8 90 10 86 40 1 98 4 03 6 01 8 85 10 80 45 1 97 4 04 6 02 8 82 10 73 50 1 97 4 05 6 04 8 78 10 67 55 1 98 4 06 6 06 8 75 10 61 60 1 98 4 08 6 09 8 72 10 55 65 1 98 4 10 6 11 8 70 10 49 70 1 99 4 12 6 13 8 67 10 43 75 1 99 4 14 6 15 8 64 10 38 80 2 00 4 16 6 18 8 62 10 33 85 2 00 4 18 6 21 8 60 10 28 90 2 00 4 21 6 24 8 58 10 23 95 2 00 4 24 6 27 8 56 ...
Page 146: ... 6 98 8 91 10 71 35 2 00 4 01 6 96 8 88 10 57 40 2 00 4 01 6 95 8 85 10 44 45 2 00 4 01 6 95 8 82 10 31 50 2 00 4 00 6 95 8 79 10 18 55 2 00 4 00 6 95 8 76 10 18 60 2 00 4 00 6 96 8 73 10 18 65 2 00 4 00 6 96 8 72 10 18 70 2 01 4 00 6 96 8 70 10 18 75 2 01 4 00 6 96 8 68 10 18 80 2 01 4 00 6 97 8 66 10 18 85 2 01 4 00 6 98 8 65 10 18 90 2 01 4 00 7 00 8 64 10 18 95 2 01 4 00 7 02 8 64 10 18 Values...
Page 147: ...Conditions for the specifiable buffer set All values must lie in the range pH 0 14 Maximum difference between two adjacent pH values 5 C step width of the same buffer solution pH 0 25 The values of buffer solution 1 must be lower than those of buffer solution 2 The difference between values for identical temperatures must be greater than 2 pH units Faulty entries are indicated in measuring mode by...
Page 148: ...ngs Editing the values of buffer solution 1 Edit using arrow keys press enter to confirm and proceed to next tempera ture value Enter the values for the first buffer solution in 5 C steps The difference to the next value must not exceed 0 25 pH unit Select buffer solution 2 for editing The difference between buffer solutions for iden tical temperatures must be greater than 2 pH units U1 Specifiabl...
Page 149: ...149 Buffer Set U1 Fill in your configuration data or use the table as original for copy Temperature C Buffer 1 Buffer 2 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 U1 Specifiable Buffer Set ...
Page 150: ...f they are out of range ERR xxx is displayed for 3 sec the display backlighting flashes red the respective maximum or minimum value is shown input must be repeated If a faulty parameter arrives through the interface HART an error message will be displayed ERR 100 199 the faulty parameter can be localized by pressing the info key Calibration errors If errors occur during calibration an error messag...
Page 151: ...vice program Configuration or calibration data defective completely reconfig ure and recalibrate the device ERR 97 NO MODULE INSTALLED No module Please have the module installed in the factory ERR 96 WRONG MODULE Wrong module Please have the module replaced in the factory ERR 95 SYSTEM ERROR System error Restart required If error still persists send in the device for repair ERR 01 NO SENSOR pH sen...
Page 152: ...ed Connect the sensor check the sensor cable and replace if necessary check the sensor connection adjust the parameter settings ERR 15 SENSOCHECK GLASS EL Sensocheck glass ERR 16 SENSOCHECK REF EL Sensocheck ref ERR 60 OUTPUT LOAD Load error Check the current loop deactivate unused current outputs ERR 61 OUTPUT 1 TOO LOW Output current 1 3 8 mA ERR 62 OUTPUT 1 TOO HIGH Output current 1 20 5 mA ERR...
Page 153: ...OW TOO LOW Flow too low ERR 73 FLOW TOO HIGH Flow too high ERR 100 INVALID SPAN OUT1 Span Out1 configuration error Selected span too small ERR 101 INVALID SPAN OUT2 Span Out2 configuration error Selected span too small ERR 102 FAILURE BUFFERSET U1 Configuration error Specifiable buffer set U1 ERR 105 INVALID SPAN I INPUT Configuration error Current input Error Messages ...
Page 154: ...missioning Disposal Local codes and regulations must be observed when disposing of the product Returns If required send the product in a clean condition and securely packed to your local contact See www knick de ...
Page 155: ...sponding icon blinks The Sensocheck message is also output as error message Err 15 glass electrode or Err 16 reference electrode for digital transmitters however only with InduCon sensors with SG The display backlighting turns red output current 1 is set to 22 mA when configured corre spondingly Sensocheck can be switched off during configuration then Sensoface is also disabled Exception After a c...
Page 156: ...on Replace sensor Calibration timer Over 80 of the calibration interval has already past The calibration interval has been exceeded Sensor defect Check the sensor and its connections see also Error Messages Err 15 and Err 16 Response time Sensor response time has increased The sensor should be replaced soon To achieve an improvement clean the sensor and soak it in buffer Sensor response time signi...
Page 157: ...ANGE SENSOR DLI Replace sensor AUTOCLAVE CYCLES OVERRUN Maximally permitted number of auto claving cycles has been reached Replace sensor or increment autoclaving counter SIP CYCLES OVERRUN Maximally permitted number of sterilizing cycles has been reached Replace sensor or increment SIP counter CIP CYCLES OVERRUN Maximally permitted number of cleaning cycles has been reached Replace sensor or incr...
Page 158: ...ART Typical Applications 8 9 17 out out 1 out 2 A201 4 20mA HART Terminal 8 9 17 out out 1 out 2 Power supply A201 Uout min 19 V 250 Ω 4 20mA HART Terminal 1 2 Repeater power supply 5 6 HART Terminal SW A001 ...
Page 159: ... Access to the device functions is regulated and limited by individually adjustable codes Passcodes see SERVICE This prevents unauthor ized modification of device settings or manipulation of the measure ment results Appropriate use of these passcodes makes them suitable as electronic signature Audit Trail Every manual change of device settings can be automatically docu mented Each change is tagged...
Page 160: ...inds you to service the sensor The TTM interval is either read automatically from the sensor settings or can be specified manually Stressing influences temperature measure ment in extreme ranges shorten the timer interval Asymmetry potential The voltage which a pH sensor provides at a pH of 7 The asymmetry potential is different for each sensor and changes with age and wear Buffer set Contains sel...
Page 161: ...utions used during calibration CIP Cleaning In Place CIP cycles are used for cleaning the process wetted parts in the process They are performed for biotech ap plications for example Depending on the application one or more chemicals are used at temperatures above 70 C This extremely stresses the sensors Digital sensors can release a message after preset number of CIP cycles This allows replacing ...
Page 162: ...ss electrode This volt age is led to the pH input of the device and is processed further as usual The adapter is directly supplied from the device ISM Intelligent Sensor Management ISM sensors have an electronic datasheet which allows the storage of additional oper ating parameters such as calibration date and settings directly in the sensor One point calibration Calibration with which only the as...
Page 163: ...e electrodes The resulting infor mation is indicated by the Sensoface smileys Sensocheck can be switched off Sensoface Provides information on the sensor condi tion The zero point slope and response time are evaluated In addition the Sensocheck information is indicated SIP Sterilization In Place CIP cycles are used for sterilizing the process wetted parts in the process They are performed for biot...
Page 164: ...r can be specified manually After expiration of the interval the adaptive maintenance timer issues a Sensoface mes sage to remind you to service the sensor Stressing influences temperature measure ment in extreme ranges shorten the timer interval Two point calibration Calibration with which the asymmetry potential zero point and slope are deter mined Two buffer solutions are required for two point...
Page 165: ...timer description 61 Adaptive maintenance timer reset 122 Alarm 40 Alarm and HOLD messages 41 Alarm setting a delay time 88 Alarm settings CONTROL input 90 Ambulance TAN 123 Assembly 13 Asymmetry potential 109 Audit Trail 159 Autoclaving counter configuration 66 Autoclaving counter description 67 Autoclaving counter error message 157 Autoclaving counter increment 122 Automatic calibration Calimati...
Page 166: ...ation timer 57 Calibration timer Sensoface 156 Calibration zero adjustment 103 Calimatic 104 CIP cleaning cycles ISM sensor 65 CIP cleaning cycles standard or ISFET sensor 63 CIP error message 157 Cleaning cycles ISM sensor configuration 64 Cleaning cycles standard or ISFET sensors configuration 62 Commissioning 7 Configuration 45 Configuration alarm 88 Configuration CIP SIP settings 64 Configurat...
Page 167: ...41 Control inputs 11 Converting slope to mV 109 Current output 1 configuring 70 Current output 2 configuring 78 D Data entry of premeasured sensors 108 Data logger explanation 10 Date and time configuration 92 Date and time usage 93 Date display 115 Declaration of Conformity 3 Decommissioning 154 Device self test 118 Device type display 120 Diagnostics 116 Diagnostics calibration data 117 Diagnost...
Page 168: ...ring values 34 Error handling 150 Error messages 151 EU Declaration of Conformity 3 Extended logbook Audit Trail 159 Extended logbook diagnostics 119 External temp measurement via current input 82 F FACE Sensoface alert 22 mA output current 75 FAIL error message 22 mA output current 75 FDA 21 CFR Part 11 159 FLASH test 118 Floating relay output 86 FLOW 87 Flow measurement alarm 90 Flow measurement...
Page 169: ... configure adaptive maintenance timer 60 ISM sensors configure autoclaving counter 66 K Keypad 31 L Linear temperature compensation 81 Logbook diagnostics 119 M Maintenance 125 Manual calibration with buffer entry 106 Manual switchover of parameter sets A B 44 Measured values display 120 Measuring mode 115 Measuring mode overview 33 Measuring points TAG GROUP 95 Memosens configuring the device 97 ...
Page 170: ...e FACE 75 Output current range 1 configuring 70 Output current range 2 configuring 78 Output current specify value 122 Output filter 72 Output signal during HOLD 39 Output signal during HOLD configuration 77 Overview device properties 9 Overview of configuration 45 Overview terminal assignments 12 P Package contents 13 Package contents documentation 3 Panel mounting 16 Parameter error 150 Paramete...
Page 171: ...duct line 127 Protective hood 15 Q Quickstart guides 3 R RAM test 118 Rated operating conditions 133 Rating plates 18 Redox calibration 112 Repair 125 Reset to factory settings 123 Returns 154 S Safety 7 Safety guide 3 Selection menu 34 Sensocheck 155 Sensocheck configuration 88 Sensoface 155 Sensoface alerts 75 Sensoface troubleshooting 150 Sensor connection 20 Sensor connection assignments 19 Se...
Page 172: ...s of U1 buffer set 149 Setting the passcodes 123 Signal colors 32 36 Signal lines 19 Signal outputs 11 SIP error message 157 SIP sterilization cycles ISM sensor 65 SIP sterilization cycles standard or ISFET sensor 63 Software version display 120 Solution Ground and Memosens 96 Specifiable buffer set U1 147 Specifications 128 Sterilization cycles ISM sensor configuration 64 Supplemental directives ...
Page 173: ...re probe selection 53 Terminal assignments 18 Terminal assignments overview 12 Test report 2 2 3 Time and date configuration 92 Time and date usage 93 Time display 115 Transaction number TAN 123 TTM configure adaptive maintenance timer 60 TTM Time To Maintenance ISM sensors 60 U U1 specifiable buffer set 147 User interface 31 W Weather protector 15 Wiring 19 Wiring examples 20 Wiring power supply ...
Page 174: ...174 Index ...
Page 175: ...175 ...
Page 176: ...shed on February 18 2022 The latest documents are available for download on our website under the corresponding product description Knick Elektronische Messgeräte GmbH Co KG Headquarters Beuckestraße 22 14163 Berlin Germany Phone 49 30 80191 0 Fax 49 30 80191 200 info knick de www knick de Local Contacts www knick international com ...