6-46 Applied Operation
PLZ-4W
■
Accurately proportionating the external voltage variation and
input current variation
You can use the pre-set resistors OFS and FSC on the PLZ-4W rear panel to adjust
the current offset value and maximum current value, respectively. The adjustment
enables control that is accurately proportional to the external voltage. The adjust-
ment is valid for the speci
fi
ed range. If you change the range, make a readjustment.
• Adjustment is made by actually supplying the load current. Prepare the power
supply to be tested.
1.
Turn on the load.
2.
Set the external voltage to 0 V.
3.
Turn OFS to set the current reading of the display measurement section
to 0 A.
4.
Set the external voltage to 10 V.
5.
Turn FSC to set the current reading of the display measurement section
to 100 % of the selected range.
6.
Turn off the load to
fi
nish the adjustment.
NOTE