5-24 Basic Operation
PLZ-4W
5.12 Short Function
The short function is used to immediately set the maximum current (CC mode) or
minimum resistance (CR mode).
When the short function is enabled, the short signal output of the J1 connector is
turned on. The short signal output terminal is connected to a relay contact (30 VDC/
1 A) and can be used to drive an external relay for high current to short the load
input terminal.
For a description of the J1 connector, see “J1 and J2 Connectors” in section 6.8,
“External Control.”
• Be sure to use a dedicated driver circuit to drive the relay for high current.
Fig. 5-12
Connection example of a relay for high current
■
Procedure
Fig. 5-13
Keys used in the short function
1.
Execute the short function.
With the load turned on, press the SHORT (SHIFT+SLEW RATE) key.
A short icon appears. When in CC mode, the value is set to the maximum
current; when in CR mode, the value is set to the minimum resistance.
• The short function setting cannot be saved to the setup memory.
2.
Clear the short function.
Press the SHORT (SHIFT+SLEW RATE) key again.
The short icon clears, and the PLZ-4W returns to the original condition
before the short function was executed.
CAUTION
+
Equipment under test
PLZ-4W Series Electronic Load
Relay
Relay for high current
J1 connector on
the rear panel
–
+
–
<19>
<20>
J1
Driver
circuit
SWITCHING
ENTER
SHORT
EDIT
SLEW RATE
RUN / STOP
UVP
PAUSE
STORE
RECALL
A
B
C
LOAD
Th/TL
FREQ / DUTY
MENU
PRESET SEQ
I
LEVEL
VRANGE
RANGE
SW ON
+CV
MODE
LOAD
SET / VSET
LOCK
SHIFT
ABC
OPP / OCP
0.00
mA
0.000
V
0.000
W
CC
SLEWRATE
0.500
A
0.01
mA/
μ
s
SET
H 33A
150V
Short icon