6-42 Applied Operation
PLZ-4W
Table6-9
J1 connector pin arrangement
*1. Valid only when the front panel settings is H range.
*2.
*3.
*4. The maximum applied voltage of the photocoupler is 30 V; the maximum current is 8 mA.
Pin
No.
Signal name
Description
1
EXT R/V CONT
Can be used in CC, CR, CV, and CP modes
0 V to 10 V correspond to 0 % to 100 % of the rated current (CC mode), rated voltage
(CV mode), or rated power (CP mode).
0 V to 10 V correspond to maximum resistance to minimum resistance (CR mode)
0
Ω
to 10 k
Ω
correspond to 0 % to 100 % or 100 % to 0 % of the rated current (CC
mode), rated voltage (CV mode), or rated power (CP mode).
0
Ω
to 10 k
Ω
correspond to maximum resistance to minimum resistance or minimum
resistance to maximum resistance (CR mode)
2
IMON
Current monitor output
10 V f.s (H/L range) and 1 V f.s (M range)
3
A COM
Connected to the negative load input terminal on the rear panel.
4
SUM I MON
Used during master/slave operation. Connected to SUM I MON of the J2 connector.
5
PRL IN+
Used during master/slave operation. Connected to PRL OUT+ of the J2 connector.
6
PRL IN-
Used during master/slave operation. Connected to PRL OUT- of the J2 connector.
7
LOAD ON/OFF CONT
Turns on the load with low (or high) TTL level signal
Pulled up the internal circuit to 5 V using 10 k
Ω
.
8
RANGE CONT 1
External range switch input
*1 *2
Pulled up the internal circuit to 5 V using 10 k
Ω
.
9
RANGE CONT 0
10
ALARM INPUT
Activates alarm with low TTL level signal input.
Pulled up the internal circuit to 5 V using 10 k
Ω
.
11
TRIG INPUT
When paused, clears the pause when a high level TTL signal is applied for 10
μ
s or
longer.
Pulled down the internal circuit to A COM using 100 k
Ω
.
12
A COM
Connected to the negative load input terminal on the rear panel.
13
LOAD ON STATUS
Turns on when load is on. Open collector output by a photocoupler.
*4
14
RANGE STATUS 1
Range status output. Open collector output by a photocoupler.
*4
15
RANGE STATUS 0
16
ALARM STATUS
Turns on when an alarm (OVP, OCP, OPP, OHP, REV, or UVP) is activated or when an
external alarm is applied. Open collector output by a photocoupler.
*4
17
STATUS COM
STATUS signal common for pins 13 to 16.
18
N.C.
19
SHORT SIGNAL OUT
Relay contact output (30 VDC/1 A)
20
SHORT SIGNAL OUT
RANGE CONT 0
RANGE CONT 1
H range
1
1
M range
1
0
L range
0
1
RANGE STATUS 0 RANGE STATUS 1
H range
OFF
OFF
M range
OFF
ON
L range
ON
OFF
Summary of Contents for PLZ-4W Series
Page 10: ...VIII Contents PLZ 4W ...
Page 20: ...1 10 General Information PLZ 4W ...
Page 38: ...2 18 Installation and Preparation PLZ 4W ...
Page 90: ...5 30 Basic Operation PLZ 4W ...
Page 192: ...9 10 Specifications PLZ 4W ...
Page 214: ...A 22 Appendix PLZ 4W ...