PLZ-4W
Applied Operation 6-37
When sequences cannot be executed
Sequences cannot be executed across different modes and ranges.
1.
You cannot execute a sequence if the mode or range of the chained
program is different. Check the settings.
2.
If the load is turned on and the current mode and range settings do
not match the settings of the sequence that you are trying to exe-
cute, you cannot execute the sequence. Turn off the load, and set
the sequence settings to match the current settings.
3.
If the load is turned off and the current mode and range settings do
not match the settings of the sequence that you are trying to exe-
cute, the settings of the sequence that you are trying to execute are
forcibly changed to the current settings, and the sequence is exe-
cuted.
Table 6-5
Before sequence execution
Sequence execution
Load on/off
condition
Mode and range settings that you
are trying to execute
Execution
Mode and range settings that are
executed
Load on
Matches the current settings
Possible
Same as the current settings
Do not match the current settings
Not possible
Load off
Matches the current settings
Possible
Same as the current settings
Do not match the current settings
Possible
Forcibly changed to the current
settings