5-14 Basic Operation
PLZ-4W
3.
Select the current and conductance range.
Press the RANGE key.
A current range pop-up menu appears.
Each time you press the RANGE key while the menu is displayed, the
range switches in the order L, M, and H.
When the desired range is highlighted, stop pressing the key.
Along with L, M, or H, the full scale value of the respective range is displayed.
This value varies depending on the model.
You cannot select separately the range for current and conductance.
4.
Select the voltage range.
Press the VRANGE (SHIFT+RANGE) key.
A voltage range pop-up menu appears.
Each time you press the VRANGE key, the voltage range switches
between 15 V and 150 V.
When the desired range is highlighted, stop pressing the key.
5.
Set the conductance [S].
Check that the display is in the basic setting entry condition (the characters
“SET” is highlighted).
While viewing the display, turn the rotary knob.
First, use coarse adjustment to roughly set the value, and then switch to
fi
ne
adjustment to adjust the value precisely.
When you press the SET/VSET key, the resistance calculated from the con-
ductance is indicated at the multi display.
6.
Turn on the load.
Press the LOAD key.
The LOAD LED illuminates, and the current
fl
ows.
The display shows the measured values (voltage, current, and wattage)
at the load input terminal.
7.
Change the conductance setting.
Turn the rotary knob to change the conductance setting.
The conductance cannot be set greater than the maximum value of the selected
range.
8.
Turn off the load.
Press the LOAD key.
The LOAD LED turns off, and the current is cut off.
Summary of Contents for PLZ-4W Series
Page 10: ...VIII Contents PLZ 4W ...
Page 20: ...1 10 General Information PLZ 4W ...
Page 38: ...2 18 Installation and Preparation PLZ 4W ...
Page 90: ...5 30 Basic Operation PLZ 4W ...
Page 192: ...9 10 Specifications PLZ 4W ...
Page 214: ...A 22 Appendix PLZ 4W ...