PAD-L
Precautions and Preparations for Use
2-3
2.3
Load
Note that the output may become unstable when one of the following loads is
connected.
2.3.1
When load current has peaks or is pulse-shaped
The current meter on the power supply indicates only mean values. Even when the
indicated value is less than the preset current value, therefore, a peak may exceed the
preset current value. In such a case, the power supply is instantaneously put into
constant-current operation mode, and the output voltage drops accordingly. If you
look carefully, you will see the constant current (CC) indicator light up dimly.
For such a load, a larger value should be preset for the constant current, or the current
capacity should be increased.
Constant current preset value
Meter indication value
(mean value)
Constant current preset value
Meter indication value
(mean value)
Fig. 2-1 Load current with a peak
Fig. 2-2
Pulsing load current