6-12
Maintenance
PAD-L
Current system adjustment procedure (continued)
Current limit
16. Adjust the RV53 variable resistor on the PCB A-200 so that the unit
ammeter indication is the same when the CURRENT/VOLT. LIMIT
switch is held down and released.
See Figs. 6-9, 6-10, and 6-11.
This completes the current system adjustment.
Table 6-2
Maximum current value
Type III
PAD
16-100L
PAD
35-50L
PAD
35-60L
PAD
60-35L
PAD
110-20L
PAD
250-8L
Maximum current [V]
105
51
61
36
20.5
8.2
Type IV
PAD
35-100L
PAD
60-60L
PAD
110-30L
PAD
250-15L
Maximum current [V]
105
61
30.5
15.5
Type V
PAD
35-200L
PAD
60-120L
PAD
110-60L
Maximum current [V]
205
125
62
PCB A-200
RV53 (Currnt limit ADJ)
RV49 (Maximum variable constant-currnt range ADJ)
Never touch this
variable resistor.
Fig. 6-9
RV49 and RV53 in the TYPE III model