PAD-L
Setup
1-7
1.4
Changing the input power voltage
(TYPE III models only)
The input power voltage of the TYPE III model is factory-set to 200 V. This can be
changed to 100 V by modifying the connections of the input terminals of the main
power transformer.
You can contact your Kikusui distributor/agent to change the input power voltage.
To perform this work yourself, follow the procedure given below.
WARNING
•
Changing the input power voltage requires removal of the top
covers of TYPE III models. This work must be carried out by
qualified personnel having sufficient expertise and familiarity with
the following procedures and safety precautions.
•
After the input power voltage is changed from 200 V to 100 V, the
current capacity of the supplied AC power cord will be
insufficient. The AC power cord must be replaced with a cable of
the necessary current capacity.
Input power voltage changing procedure (TYPE III models only)
1.
Confirm by inspection that the AC power cord has not been connected to
the input terminal board of the unit.
2.
Remove the top panel. See Fig. 1-2.
Black binding screws M3-6L
6 pieces
Fig. 1-2
Removing the top panel from the TYPE III model