7-2
Specifications
PAD-L
7.1
TYPE III models specifications
Unless otherwise specified, the specifications of the unit are based on the following
conditions.
•
The load is a pure resistance.
•
The - (neg.) output terminal is connected to the GND terminal.
•
The unit should be used after 30 minutes warming-up time.
Standard value and theoretical value do not guarantee performance. They should be
referred to as target values only.
Type III
PAD
16-100L
PAD
35-50L
PAD
35-60L
PAD
60-35L
PAD
110-20L
PAD
250-8L
Input
Input voltage and frequency
200/100 VAC±10 %, 50/60 Hz, 1 φ
(Changeable by wiring terminals of the main power transformer)
Power consumption, at 200VAC, rated load
Approx.
3.3 kVA
Approx.
3.3 kVA
Approx.
3.8 kVA
Approx.
3.4 kVA
Approx.
3.8 kVA
Approx.
3.4 kVA
Output
Voltage
Rated voltage
16 V
35 V
35 V
60 V
110 V
250 V
Variable range
0 to 16V
0 to 35 V
0 to 35 V
0 to 60 V
0 to 110 V
0 to 250 V
Resolution (theoretical value) *1
3 mV
6.3 mV
6.3 mV
11 mV
20 mV
45 mV
Number of turns of panel control
10 turns
Current
Rated current
100 A
50 A
60 A
35 A
20 A
8 A
Variable range
0 to 100 A
0 to 50 A
0 to 60 A
0 to 35 A
0 to 20 A
0 to 8 A
Resolution (theoretical value) *1
47 mA
140 mA
170 mA
120 mA
68 mA
27 mA
Number of turns of panel control
1 turn *4
Constant voltage characteristics
Ripple and noise (5Hz to 1MHz) RMS
0.5 mV
0.5 mV
0.5 mV
0.5 mV
1 mV
5 mV
0.005%+1mV 0.005%+1mV 0.005%+1mV 0.005%+1mV 0.005%+1mV
+2mV
0.005%+2mV 0.005%+2mV 0.005%+2mV 0.005%+2mV 0.005%+2mV
+3mV
Transient response (standard value) *2, *3
Temperature coefficient (standard value)
50 ppm/℃
Constant current characteristics
Ripple and noise (5Hz to 1MHz) RMS
100 mA
10 mA
10 mA
8 mA
4 mA
4 mA
Source effect (to±10% of AC input voltage)
3 mA
3 mA
3 mA
3 mA
1 mA
1 mA
Load effect (to 0 to 100% of output voltage)
5 mA
5 mA
5 mA
3 mA
3 mA
3 mA
Constant voltage operation indication
C.V, green lamp indication
Constant current operation indication
C.C, red lamp indication
*1: The value is calculated from the number of turns of the wire to wound potentiometer. In practice, use 3 to 5 times each value as a target.
*2: Measured using the remote sensing function.
*3: Time necessary for output voltage to return to ±(0.05%+10mV) max. of rated value at 5 to 100% changes of output current.
*4: 1 turn, both of coarse and fine adjustment
1 turn
50 μs
Source effect
(to±10% of AC input voltage) *2
Load effect
(to 0 to 100% of output current) *2