7-6
KHA1000
7.2.2
Auto Range, Voltage Range, and Current Range
Sub-Menu
Basic Measurement
Sub Menu
The settings are also applied to those in Sections 7.3
” and
“
.” Note, however, that Auto Range is not applied.
The peak current that can be measured is four times the range value in the 0.5 to 10 A
range and 2.5 times in the 20 A range.
•
If waveforms with a large crest factor are input while Auto Range is ON, the
range may not remain constant. In this case, set Auto Range to OFF.
•
The maximum value of input current is 50 Apeak. Exceeding this value
may burn the current detector.
•
If the current detector overheats, the OHP icon appears on the upper part
of the display. Immediately shut down the power to the EUT to cut the
input current of this product. Restart the test after the OHP icon
disappears.
F 1
F 1
F 2
F 2
F 3
F 3
F 4
F 4
F 5
F 5
F 6
F 6
Page
2/3
F1
Auto Range
OFF
ON
Select ON to enable Auto Range for voltage and
current.
F2
Voltage Range
150 V
300 V
Select the voltage range. The scale sensitivity is
displayed on the upper part of the graph according to
the selected range.
F3
Current Range
0.5 A
1 A
2 A
5 A
10 A
20 A
Select the current range. Select it according to the
input current of the EUT. The scale sensitivity is
displayed on the upper part of the graph according to
the selected range.
CAUTION
Summary of Contents for KHA1000
Page 1: ...Part No Z1 003 482 I0043901 Oct 2021 Operation Manual Harmonic Flicker Analyzer KHA1000 ...
Page 8: ...vi KHA1000 ...
Page 20: ...xviii KHA1000 Rear panel 1 2 3 4 5 6 7 8 9 10 11 12 13 14 ...
Page 22: ...xx KHA1000 ...
Page 23: ...Overview This chapter provides an overview of the product and explains its features 1 ...
Page 30: ...1 8 KHA1000 ...
Page 68: ...3 20 KHA1000 ...
Page 69: ...Basic Operation This chapter explains basic operation methods 4 ...
Page 156: ...5 54 KHA1000 ...
Page 260: ...9 6 KHA1000 ...
Page 261: ...Specifications This chapter explains the specifications of this product 10 ...
Page 272: ...I 4 KHA1000 ...
Page 273: ......