5-32
KHA1000
View Setting
Sub Menu
A different sub-menu is displayed for each view type selected. The sub-menus of a
results list are given below.
Results list
•
Items (100 % and margin) are displayed in JIS 2003/Ed1.0. In other standards,
they are displayed in a pale color. Standard judgment is made at the ratio of time
(accumulated value in entire test period) in which the selected value (100 % or
margin) is exceeded, against the entire test period.
Explanation of display items
Page 5-34
F1
Item
100 %
Margin
Limit value Value set as
margin
F2
Comment Input
Sub Menu
Memo
Sub Menu
Model name
Sub Menu
Type
Sub Menu
Serial No.
Sub Menu
F5
Scroll
Used to view a hidden part. Turn the small knob to
vertically scroll over the display. This is used when the
display is divided into two or four windows. In the
1-screen, it is displayed in a pale color.
See
Display item
Description
No.
Harmonic order
Max(A)
Maximum value (A) during test period
•
Part in which the maximum value exceeds 150 % of the limit value is
displayed in red.
•
This differs from the harmonics list (part in which 100 % of the limit
value is exceeded is displayed in red).
Ave(A)
Average value (A) during entire test period
•
Part in which the average value in the entire test period exceeds 100 %
of the limit value is displayed in red.
100 %
or
margin %
100 % is the ratio (%) of time in which limit values (accumulated value of
entire test period) are exceeded, against the entire measurement time.
Margin % is displayed as follows when the limit value replaces the set
value of the margin:
•
The part in which the ratio (%) of time (accumulated value of entire test
period) in which measured values exceed 100 % of the limit value up to
150 % of that limit value, against the entire test period, exceeds 10 % is
displayed in red.
Judge
Judgment of each harmonic order (PASS/FAIL/WARN)
•
WARN is displayed when the margin is exceeded.
“N/A” is displayed when limit values are not applied in the following cases:
•
When there is not limit value
•
When the measured value of harmonic current is smaller than 0.6 % of
input current or 5 mA, whichever is larger.
Margin---%
Margin to limit value
•
Displayed in 1-screen view only
Result
Final test result (PASS/FAIL/WARN)
•
WARN is displayed when the margin is exceeded.
POHC
Maximum value of POHC in test period
•
Displayed in red when POHC exceeds POHCLim
POHCLim
POHC calculated from limit value applied
Lim2:200 %
Displayed when 200 % of the limit value is applied.
Summary of Contents for KHA1000
Page 1: ...Part No Z1 003 482 I0043901 Oct 2021 Operation Manual Harmonic Flicker Analyzer KHA1000 ...
Page 8: ...vi KHA1000 ...
Page 20: ...xviii KHA1000 Rear panel 1 2 3 4 5 6 7 8 9 10 11 12 13 14 ...
Page 22: ...xx KHA1000 ...
Page 23: ...Overview This chapter provides an overview of the product and explains its features 1 ...
Page 30: ...1 8 KHA1000 ...
Page 68: ...3 20 KHA1000 ...
Page 69: ...Basic Operation This chapter explains basic operation methods 4 ...
Page 156: ...5 54 KHA1000 ...
Page 260: ...9 6 KHA1000 ...
Page 261: ...Specifications This chapter explains the specifications of this product 10 ...
Page 272: ...I 4 KHA1000 ...
Page 273: ......