8-24
KHA1000
“IEC
:ED30:ED10
”
IEC Ed3.0/Ed1.0
“JIS
:2011:ED10
”
JIS 2019/Ed1.0
“JIS
:2011:ED20
”
JIS 2019/Ed2.0
“JIS:2019:ED10”
JIS 2019/Ed1.0
“JIS:2019:ED20”
JIS 2019/Ed2.0
“IEC:ED50:ED10”
IEC Ed5.0/Ed1.0
“IEC:ED50:ED21”
IEC Ed5.0/Ed2.1 (Default)
When *RST is sent, settings are made as shown in Table 8-4.
Response
Returns a test standard.
• The conventional setting value “IEC:ED30A2:ED20A1” and “IEC:ED30A2:ED10”
can be used, but the IEC 61000-3-2 Ed5.0 standard is applied.
• The conventional setting value “JIS:2011:ED10” and “JIS:2011:ED20” can be
used, but the JIS C61000-3-2(2019) standard is applied.
HARM:CLAS
Sets the class of EUT.
The reference value for determining conformance to the standard is the limit value
corresponding to the class.
Command
HARMonic:CLASs ”{A|B|C|D}”
HARMonic:CLASs?
Parameter
Value:
“A”
Class A (Default)
“B”
Class B
“C”
Class C
“D”
Class D
When *RST is sent, settings are made as shown in Table 8-4.
Response
Returns the class of EUT.
HARM:FREQ:NOM
Sets the nominal frequency. Set the nominal frequency according to the rating of the EUT.
Executing this command simultaneously sets VF:FREQeuncy:NOMinal to the same
value.
Command
HARMonic:FREQuency:NOMinal {50|60}
HARMonic:FREQuency:NOMinal?
Parameter
Value:
50 or 60 (The default is 50.)
Unit:
Hz
Summary of Contents for KHA1000
Page 1: ...Part No Z1 003 482 I0043901 Oct 2021 Operation Manual Harmonic Flicker Analyzer KHA1000 ...
Page 8: ...vi KHA1000 ...
Page 20: ...xviii KHA1000 Rear panel 1 2 3 4 5 6 7 8 9 10 11 12 13 14 ...
Page 22: ...xx KHA1000 ...
Page 23: ...Overview This chapter provides an overview of the product and explains its features 1 ...
Page 30: ...1 8 KHA1000 ...
Page 68: ...3 20 KHA1000 ...
Page 69: ...Basic Operation This chapter explains basic operation methods 4 ...
Page 156: ...5 54 KHA1000 ...
Page 260: ...9 6 KHA1000 ...
Page 261: ...Specifications This chapter explains the specifications of this product 10 ...
Page 272: ...I 4 KHA1000 ...
Page 273: ......