4-20
KHA1000
•
To prevent an electric shock, do not touch the SOURCE and LOAD
terminals of this product.
•
Do not touch the OUTPUT terminal of the AC Power Supply.
•
Do not touch the INPUT and OUTPUT terminals of the Line
Impedance Network.
4.5.1
AC Pow. OUTPUT, AC Pow. Output Voltage,
AC Pow. Output Frequency, and AC Pow. Voltage Range
Menu
This menu turns on/off the AC Power Supply OUTPUT.
In the ON state, the icon in the upper part of the display shows a turned-on light
bulb.
This menu sets the AC Power Supply output voltage. The setting range is from 0.0
V to 305.0 V. It corresponds to the F4 AC Pow. Voltage Range. The setting should
be in accordance with the power rating of the EUT.
This menu sets the AC Power Supply output frequency. Select 50 Hz or 60 Hz
according to the power rating of the EUT.
This menu sets the AC Power Supply output voltage range.
The 100 V range is used when the output voltage setting range is from 0 V to 152.5 V.
The 200 V range is used when the output voltage setting range is from 0 V to 305.0 V.
F1
AC Pow.
OUTPUT
ON
OFF
WARNING
F 1
F 1
F 2
F 2
F 3
F 3
F 4
F 4
F 5
F 5
F 6
F 6
Page
1/1
See
F2
AC Pow.
Output Volts
numeric
value
F3
AC Pow.
Output Freq
50 Hz
60 Hz
F4
AC Pow.
Voltage Range
100 V
200 V
Summary of Contents for KHA1000
Page 1: ...Part No Z1 003 482 I0043901 Oct 2021 Operation Manual Harmonic Flicker Analyzer KHA1000 ...
Page 8: ...vi KHA1000 ...
Page 20: ...xviii KHA1000 Rear panel 1 2 3 4 5 6 7 8 9 10 11 12 13 14 ...
Page 22: ...xx KHA1000 ...
Page 23: ...Overview This chapter provides an overview of the product and explains its features 1 ...
Page 30: ...1 8 KHA1000 ...
Page 68: ...3 20 KHA1000 ...
Page 69: ...Basic Operation This chapter explains basic operation methods 4 ...
Page 156: ...5 54 KHA1000 ...
Page 260: ...9 6 KHA1000 ...
Page 261: ...Specifications This chapter explains the specifications of this product 10 ...
Page 272: ...I 4 KHA1000 ...
Page 273: ......