2-12
KHA1000
Fig.2-7
Voltage input sensing, and wires used
Use wires UL1015 and AWG18 with a conductor exposure margin of 10 mm (indicated on the rear
panel). To secure the wires, tighten the screw part of the terminal with the accessory screwdriver.
2.4.4
Connecting the AC Power Supply or Line Impedance
Network (to SOURCE Terminals)
Connect the AC power supply or line impedance network to the SOURCE
terminals. Match the polarities (L, N, and G) of the terminals and the output
terminal polarities of the AC power supply or line impedance network to each other
(G is a protective grounding or grounding terminal).
Using the terminal covers
To prevent an electric shock, covers are attached to the LOAD and SOURCE
terminals.
Wire to be used and terminal treatment
Table 2-4
Select a wire according to the input current of the EUT. Attach a cramping terminal
to terminate the wire. Use a cramping terminal that has a screw mounting part with a
hole diameter of 6 mm and fits the wire.
INPUT
OUTPUT
L
N
G
EUT
AC power supply
AC power supply
G N L
G
N
L
G N L
Voltage sensing terminal
Enlarged view of plug
V
±
N
L
V
±
N
L
Wiring at time
of shipment
Terminal screw part
See
See
Summary of Contents for KHA1000
Page 1: ...Part No Z1 003 482 I0043901 Oct 2021 Operation Manual Harmonic Flicker Analyzer KHA1000 ...
Page 8: ...vi KHA1000 ...
Page 20: ...xviii KHA1000 Rear panel 1 2 3 4 5 6 7 8 9 10 11 12 13 14 ...
Page 22: ...xx KHA1000 ...
Page 23: ...Overview This chapter provides an overview of the product and explains its features 1 ...
Page 30: ...1 8 KHA1000 ...
Page 68: ...3 20 KHA1000 ...
Page 69: ...Basic Operation This chapter explains basic operation methods 4 ...
Page 156: ...5 54 KHA1000 ...
Page 260: ...9 6 KHA1000 ...
Page 261: ...Specifications This chapter explains the specifications of this product 10 ...
Page 272: ...I 4 KHA1000 ...
Page 273: ......