HARM:LIN:STAT
Sets whether to use the reference impedance.
Command
HARMonic:LIN:STATe (ON|OFF|1|0)
HARMonic:LIN:STATe?
Parameter
Value:
OFF(0)
The reference impedance is not used (default).
ON(1)
The reference impedance is used.
Response
Returns whether the reference impedance is being used in <NR1> format.
8.7.3
Voltage Fluctuation Measurement Mode
VF:STAN:LIM
Sets the limitation standard.
Command
VF:STANdard:LIMitation[:NAME] "IEC 61000-3-3 Ed2.0"
VF:STANdard:LIMitation[:NAME]?
Parameter
Value:
"IEC 61000-3-3 Ed2.0" (Default)
When *RST is sent, settings are made as shown in Table 8-4.
Response
Returns the limitation standard.
•
To perform a test based on the IEC 61000-3-3 Ed3.1 standard, set the
VF:STAN:MTEC command parameter to “IEC 61000-4-15 Ed2.0.” When the
standard for measurement techniques is set to IEC 61000-4-15 Ed2.0, the
standard is automatically set to IEC 61000-3-3 Ed3.1 in test conditions, on the
KHA1000 display, and in reports. The measuring instrument requirements of the
IEC 61000-3-3 Ed3.1 standard are specified in IEC 61000-4-15 Ed2.0.
Summary of Contents for KHA1000
Page 1: ...Part No Z1 003 482 I0043901 Oct 2021 Operation Manual Harmonic Flicker Analyzer KHA1000 ...
Page 8: ...vi KHA1000 ...
Page 20: ...xviii KHA1000 Rear panel 1 2 3 4 5 6 7 8 9 10 11 12 13 14 ...
Page 22: ...xx KHA1000 ...
Page 23: ...Overview This chapter provides an overview of the product and explains its features 1 ...
Page 30: ...1 8 KHA1000 ...
Page 68: ...3 20 KHA1000 ...
Page 69: ...Basic Operation This chapter explains basic operation methods 4 ...
Page 156: ...5 54 KHA1000 ...
Page 260: ...9 6 KHA1000 ...
Page 261: ...Specifications This chapter explains the specifications of this product 10 ...
Page 272: ...I 4 KHA1000 ...
Page 273: ......