8-8
KHA1000
8.4.3
USB Interface
Setting the USB
1.
Press the SYSTEM key.
The System Setting display appears.
2.
Press the F2 key (I/F Select) a few times or turn the small knob to select
USB.
3.
Turn off the POWER switch and turn it on again.
The I/F selection is set.
Service Request
Service Request and Serial Polling functions are installed.
USB function
Compliant to USB Specification 2.0
Compliant to USBTMC Specification 1.0 and USBTMC-USB488 Specification 1.0
Communication speed: 12 Mbps (Full Speed)
VID (Vendor ID): 0x0B3E
PID (Product ID): 0x100D
•
To control this product through the USB interface, a device driver corresponding
to the USB T&M class (USBTMC) is required. The USBTMC drive is
automatically installed by one of the following VISA libraries:
•
KI-VISA 3.0.4 or later
(Can be downloaded from our web site)
•
NI-VISA 3.3.0 or later
(Can be downloaded from the web site of National Instruments)
•
Keysight VISA (Keysight IO Libraries) Suite 14.1 or later
(Can be downloaded from the web site of Keysight Technologies)
See
Summary of Contents for KHA1000
Page 1: ...Part No Z1 003 482 I0043901 Oct 2021 Operation Manual Harmonic Flicker Analyzer KHA1000 ...
Page 8: ...vi KHA1000 ...
Page 20: ...xviii KHA1000 Rear panel 1 2 3 4 5 6 7 8 9 10 11 12 13 14 ...
Page 22: ...xx KHA1000 ...
Page 23: ...Overview This chapter provides an overview of the product and explains its features 1 ...
Page 30: ...1 8 KHA1000 ...
Page 68: ...3 20 KHA1000 ...
Page 69: ...Basic Operation This chapter explains basic operation methods 4 ...
Page 156: ...5 54 KHA1000 ...
Page 260: ...9 6 KHA1000 ...
Page 261: ...Specifications This chapter explains the specifications of this product 10 ...
Page 272: ...I 4 KHA1000 ...
Page 273: ......