8-30
KHA1000
HARM:SMO
Sets the type of smoothing under test standard JIS 2003/Ed1.0.
Command
HARMonic:SMOothing {NOME|TCONst|AVERage}
HARMonic:SMOothing?
Parameter
Value:
NONE
No (Default)
TCONst
Time constant 1.5 s
AVERage
Average value
When *RST is sent, settings are made as shown in Table 8-4.
Response
Returns the type of smoothing.
HARM:TOBS
Sets the type of equipment operation for setting the observation period (measurement time).
Command
HARMonic:TOBServation
{QSTationary|SCYClic|RANDom|LCYClic}
HARMonic:TOBServation?
Parameter
Value:
QSTationary
Quasi stationary (Default)
SCYClic
Short cyclic
RANDom
Random
LCYClic
Long cyclic
When *RST is sent, settings are made as shown in Table 8-4.
Response
Returns the type of equipment operation for setting the observation period
(measurement time).
HARM:TTIM
Sets the measurement time.
Command
HARMonic:TTIMe{<numeric>|MIN|MAX}
HARMonic:TTIMe?
Parameter
Value:
1 to 9600 (The default is 150.)
Value:
1 to 150 (Under test standard JIS:2003:ED10)
Unit:
S
When *RST is sent, settings are made as shown in Table 8-4.
Response
Returns the measurement time in the <NR3> format.
Summary of Contents for KHA1000
Page 1: ...Part No Z1 003 482 I0043901 Oct 2021 Operation Manual Harmonic Flicker Analyzer KHA1000 ...
Page 8: ...vi KHA1000 ...
Page 20: ...xviii KHA1000 Rear panel 1 2 3 4 5 6 7 8 9 10 11 12 13 14 ...
Page 22: ...xx KHA1000 ...
Page 23: ...Overview This chapter provides an overview of the product and explains its features 1 ...
Page 30: ...1 8 KHA1000 ...
Page 68: ...3 20 KHA1000 ...
Page 69: ...Basic Operation This chapter explains basic operation methods 4 ...
Page 156: ...5 54 KHA1000 ...
Page 260: ...9 6 KHA1000 ...
Page 261: ...Specifications This chapter explains the specifications of this product 10 ...
Page 272: ...I 4 KHA1000 ...
Page 273: ......