5-18
KHA1000
5.3.2
Measuring Time, Ignore Over 19th If Dropping,
Smoothing, and Wave Check/Envelope
Menu
Set the measuring time. The setting range is 1 to 150 seconds. Set this value, taking
into consideration the operation cycle of the EUT.
If harmonic current above the 19th is dropping slightly, up to the 19th is to be
judged.
Use this menu to check the waveform of Class D. It displays the ratio (%) that is
included in the envelope for judgment. It also displays other voltage values and the
peak current of the current value in both the positive and negative direction.
F 1
F 1
F 2
F 2
F 3
F 3
F 4
F 4
F 5
F 5
F 6
F 6
Page
2/3
F2
Measuring Time
(s)
Numeric
value
F3
Ignore > 19th
gentle fall
Enabled
Disabled
F4
Smoothing
None
1.5 s
Average 4
None
Select when harmonic current is stable.
1.5 s
Select when harmonic current is fluctuating. This is the most general
method of using 1.5 second smoothing filter.
Average 4
Select when harmonic current is fluctuating. This is a simplified
smoothing function.
F5
Wave Check/
Envelope
Sub Menu
Press
Summary of Contents for KHA1000
Page 1: ...Part No Z1 003 482 I0043901 Oct 2021 Operation Manual Harmonic Flicker Analyzer KHA1000 ...
Page 8: ...vi KHA1000 ...
Page 20: ...xviii KHA1000 Rear panel 1 2 3 4 5 6 7 8 9 10 11 12 13 14 ...
Page 22: ...xx KHA1000 ...
Page 23: ...Overview This chapter provides an overview of the product and explains its features 1 ...
Page 30: ...1 8 KHA1000 ...
Page 68: ...3 20 KHA1000 ...
Page 69: ...Basic Operation This chapter explains basic operation methods 4 ...
Page 156: ...5 54 KHA1000 ...
Page 260: ...9 6 KHA1000 ...
Page 261: ...Specifications This chapter explains the specifications of this product 10 ...
Page 272: ...I 4 KHA1000 ...
Page 273: ......