Keysight N9038A MXE Service Guide 99
Instrument Messages
Event Messages
–224
Illegal parameter
value; Illegal
identifier
<identifier>. This
value may already be
in use.
The value for the LXI LAN identifier parameter must be unique (that is, LAN0
and LAN7 must have different identifier strings)
–224
Illegal parameter
value; Index out of
range
When querying the LXI Event Log or the Servo Log, an index may be used to
look at a specific entry. This error occurs if the index provided does not point to
a valid entry
–224
Illegal parameter
value; Invalid list
length
You are trying to set some list measurement settings, but the multiple lists that
you sent had differing lengths. The number of settings must be consistent from
list to list
–224
Illegal parameter
value; LXI Event
<event> already
exists.
This error occurs when you try to add an LXI Event that has already been added
–224
Illegal parameter
value; LXI Event
<event> contains
illegal characters.
When a new LXI Event is created, it may not use the comma, semicolon, or
newline characters. All other printable ASCII characters are valid
–224
Illegal parameter
value; LXI Event
<event> does not
exist.
The requested event has not been added yet
–224
Illegal parameter
value; Measurement
not available
You tried to turn on a measurement that is not available in the current mode
–224
Illegal parameter
value;This instrument
is always DC coupled
You cannot set AC coupling in this instrument
–224
Illegal parameter
value;This model is
always AC coupled
You cannot set DC coupling in this analyzer
–225
Out of memory
There is not enough memory to perform the requested operation
–225
Out of memory;
Insufficient
resources to load
Mode <mode name>
If you attempt to load a mode via SCPI that will exceed memory capacity, the
Mode does not load and this message is returned. “mode name” is the SCPI
parameter for the Mode in question (for example, SA for Spectrum Analyzer
Mode.) You can free up resources using the
System, Power On,
Configure Applications
menu
–225
Out of memory; Memory
limit caused Data
Acquisition to be
truncated
Err#
Message
Verbose/Correction Information
Summary of Contents for N9038A
Page 4: ...4 ...
Page 16: ...16 Contents ...
Page 130: ...130 Keysight N9038A MXE Service Guide Instrument Messages Condition Messages ...
Page 354: ...354 Keysight N9038A MXE Service Guide Analog Digital IF A3 Digital I F Troubleshooting ...
Page 406: ...406 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Page 429: ...448 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Reference Block Diagram ...
Page 431: ...450 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams ...
Page 627: ...646 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Page 639: ...658 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Page 643: ...662 Keysight N9038A MXE Service Guide Instrument Software Software Updates ...
Page 741: ...760 Keysight N9038A MXE Service Guide Tests Intermittent Unsteady Drifting Disturbances ...