686 Keysight N9038A MXE Service Guide
Tests
Amplitude Accuracy - Preamp On
Amplitude Accuracy - Preamp On
Test Limits
See test limits in the test result
.
Overview
A signal generator is used as the signal source for the test. The source
amplitude is varied using the signal source amplitude control. The attenuation
and resolution bandwidth are varied on the EMI receiver. The source amplitude
is measured by the power meter and receiver at each setting, and the values
compared. The difference between each pair of measurements indicates the
amplitude accuracy.
This measurement is performed for both RF inputs.
Table 22-13
Absolute Amplitude Accuracy - Preamp On - Required Equipment
Item
Critical Specifications
Recommended
Keysight Model
Adapter
Type-N (m), to 3.5 mm (m)
Frequency: 10 MHz to 18 GHz
VSWR: < 1.1:1
1250-1743
Adapter
3.5 mm (f) to 3.5 mm (f)
Frequency: 10 MHz to 26.5 GHz
VSWR: < 1.1:1
83059B
Adapter
a
2.4 mm (f) to 3.5 mm (m)
Frequency: 10 MHz to 26.5 GHz
VSWR: < 1.1:1
11901D
Attenuator, 30 dB
Type-N (m) to Type-N (f)
Accuracy: < 0.5 dB at 50 MHz
11708A
Attenuator, 30 dB
b
3.5 mm (m) to 3.5 mm (f)
Accuracy: < 0.5 dB at 50 MHz
8493C
Option 030
Cable 1 meter
3.5 mm (m) to 3.5 mm (m)
Frequency: 10 MHz to 1.515 GHz
VSWR: < 1.4:1
11500E
Cable 120 cm
BNC (m) to BNC (m)
Frequency: 10 MHz
10503A
Power Meter
Compatible with power sensor
E4418B
Power Sensor
Frequency Range: 50 MHz
Amplitude Range:
−
25 dBm to 10 dBm
8485A
Power Splitter
3.5 mm (f)
Frequency Range: 50 MHz to 26.5 GHz
Tracking between ports: < 0.25 dB
11667B
Signal Generator
Typical Temperature Stability: 0.01 dBc/
°
C
PSG
a. Only required for Option 544 (44 GHz Frequency Range)
b. Only required for Option C35 (3.5 mm RF Input 1) or Option 544 (44 GHz Frequency Range)
Summary of Contents for N9038A
Page 4: ...4 ...
Page 16: ...16 Contents ...
Page 130: ...130 Keysight N9038A MXE Service Guide Instrument Messages Condition Messages ...
Page 354: ...354 Keysight N9038A MXE Service Guide Analog Digital IF A3 Digital I F Troubleshooting ...
Page 406: ...406 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Page 429: ...448 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Reference Block Diagram ...
Page 431: ...450 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams ...
Page 627: ...646 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Page 639: ...658 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Page 643: ...662 Keysight N9038A MXE Service Guide Instrument Software Software Updates ...
Page 741: ...760 Keysight N9038A MXE Service Guide Tests Intermittent Unsteady Drifting Disturbances ...