Keysight N9038A MXE Service Guide 621
Assembly Replacement Procedures
Input Connector Assembly
Replacement
1.
. Place the Input Connector Assembly into position in
the chassis. Replace the two screws
(1)
(0515-0372) to attach the Input
Connector Assembly to the chassis. Torque to 9 inch-pounds.
2.
Reconnect the semi-rigid cable W25 to the Input Connector Assembly.
Torque to 10 inch-pounds.
3.
Replace the Front Frame Assembly. Refer to the
replacement procedure.
4.
Replace the instrument outer case. Refer to the
replacement procedure.