Keysight N9038A MXE Service Guide 177
RF Preselector Section
RF Preselector Section Troubleshooting
Figure 5-5
A21 RF Preselector Input Assembly Connections - Bottom Side
Verifying Control for The SW1 Cal Switch And The SW2 Transfer
Switch
1.
Remove the instrument dress cover (MP24) and the top brace (MP10).
Refer to
Chapter 18, “Assembly Replacement Procedures”
for instructions
on removing these covers.
2.
, remove W2 from A21 RF Preselector Input
assembly.
3.
Turn the instrument on and allow it to complete its boot up process,
ignoring any error messages.
4.
Turn the instrument self-alignment routine off by pressing
System
,
Alignments
,
Auto Align
,
Off
.
SW1 Cal Switch Control
5.
Switch the instrument input to RF Input 2 by pressing
Input / Output
,
RF
Input
,
RF Input Port
,
RF Input 2
.
6.
Verify that the RF Calibrator signal is turned off by pressing
Input /
Output
,
RF Calibrator
,
Off
.
7.
Measure the DC voltages on pins 11 through 20 on P600 of the A21 RF
Preselector Input assembly, as shown in
, and verify that they
are what is expected for the RF Calibrator Off as shown in
.
If the voltage levels are incorrect the most likely cause of the problem is
the A21 RF Preselector Input assembly itself.
Summary of Contents for N9038A
Page 4: ...4 ...
Page 16: ...16 Contents ...
Page 130: ...130 Keysight N9038A MXE Service Guide Instrument Messages Condition Messages ...
Page 354: ...354 Keysight N9038A MXE Service Guide Analog Digital IF A3 Digital I F Troubleshooting ...
Page 406: ...406 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Page 429: ...448 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Reference Block Diagram ...
Page 431: ...450 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams ...
Page 627: ...646 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Page 639: ...658 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Page 643: ...662 Keysight N9038A MXE Service Guide Instrument Software Software Updates ...
Page 741: ...760 Keysight N9038A MXE Service Guide Tests Intermittent Unsteady Drifting Disturbances ...