Keysight N9038A MXE Service Guide 703
Tests
Scale Fidelity
Figure 22-8
Scale Fidelity Setup
Procedure
1.
Configure the equipment as shown in
.
2.
If the auto alignment for the analyzer has not been performed within the
past 24 hours, press
System
,
Alignments
,
Align Now
,
All
to perform the
auto alignment routine.
3.
On the signal generator, press
PRESET
, then set the controls as follows:
Frequency
,
50
,
MHz
Amplitude
,
+5
,
dBm
RF
,
On
4.
Select RF Input 1 on the receiver by pressing
Input/Output
,
RF Input
,
RF
Input Port
,
RF Input
.
5.
Preset the receiver by pressing
Mode
,
Spectrum Analyzer,
Mode Preset
.
6.
Set up the receiver by pressing:
Mode Setup
,
EMC Standard
,
None
Input/Output
,
More
,
Freq Ref In
,
External
FREQ Channel, Center Freq
,
50
,
MHz
SPAN X Scale, Span
,
1, MHz
AMPTD Y Scale,
Ref Level
,
0
,
dBm
Meas Setup, Average/Hold Number, 10
,
Enter
Summary of Contents for N9038A
Page 4: ...4 ...
Page 16: ...16 Contents ...
Page 130: ...130 Keysight N9038A MXE Service Guide Instrument Messages Condition Messages ...
Page 354: ...354 Keysight N9038A MXE Service Guide Analog Digital IF A3 Digital I F Troubleshooting ...
Page 406: ...406 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Page 429: ...448 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Reference Block Diagram ...
Page 431: ...450 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams ...
Page 627: ...646 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Page 639: ...658 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Page 643: ...662 Keysight N9038A MXE Service Guide Instrument Software Software Updates ...
Page 741: ...760 Keysight N9038A MXE Service Guide Tests Intermittent Unsteady Drifting Disturbances ...