Keysight N9038A MXE Service Guide 587
Assembly Replacement Procedures
Reference Assembly
Reference Assembly
Removal
1.
Remove the instrument outer case. Refer to the
removal procedure.
2.
Remove the top brace. Refer to the
Top Brace and Power Supply Bracket
removal procedure.
3.
Refer to
. Locate the Reference assembly
(6)
.
4.
Refer to
. Remove cables W41, W42, and W43 from the
reference assembly using the 5/16-inch wrench.
5.
Remove cables W18, W19, W20, W21, and W23 from the Reference
assembly
(6)
.
Figure 18-45
Reference Board Cables
6.
Remove the two screws
(1)
.
7.
Remove the Reference board by pulling up to disconnect from the
motherboard and pull out of the chassis.
Summary of Contents for N9038A
Page 4: ...4 ...
Page 16: ...16 Contents ...
Page 130: ...130 Keysight N9038A MXE Service Guide Instrument Messages Condition Messages ...
Page 354: ...354 Keysight N9038A MXE Service Guide Analog Digital IF A3 Digital I F Troubleshooting ...
Page 406: ...406 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Page 429: ...448 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Reference Block Diagram ...
Page 431: ...450 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams ...
Page 627: ...646 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Page 639: ...658 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Page 643: ...662 Keysight N9038A MXE Service Guide Instrument Software Software Updates ...
Page 741: ...760 Keysight N9038A MXE Service Guide Tests Intermittent Unsteady Drifting Disturbances ...